User’s Manual for Agera
and EasyMatch Essentials v 1.0
44
Figure 60. New Workspace
Figure 61. Name the New Workspace
System Settings: Standardization
From the
WORKSPACE
menu, select
STANDARDIZATION
. You can also press the Standardization button
in the Status bar as a shortcut. The instrument can report the current port plate and UV Mode on the
status bar. Agera can save multiple modes contains different UV settings (nominal, excluded and
calibration) and Area of View (LAV1, LAV2, and SAV).
Each Agera has a specific set of Black Glass and White Tile unique to the instrument. These are not
interchangeable with other instruments. Position the standard tiles on the instrument with the white
line matching the white line on the port plate.
•
READ BLACK GLASS
: Place the Agera black glass on the sensor port. Press
READ
to continue.
Figure 62. Read Black Glass