Service and Maintenance
ELISYS QUATTRO
User Manual HUMAN
6-7
6.5
Replacement of Mains Fuse
The mains fuse is on the back of the analyzer.
1.
Switch off the system.
2.
Disconnect the system from the mains power.
3.
Pull out the fuse carrier with a screw driver.
Figure 6-3:
Fuse carrier
4.
Change the faulty fuse(s):
Fuse: 4 A T, 250 V
Figure 6-4:
Fuse carrier with fuses
5.
Insert the fuse carrier.
6.
Connect the main power.
7.
Switch on the system.