Performing a Test Run
ELISYS QUATTRO
User Manual HUMAN
4-25
4.7
Quality Control Analysis Report (Levey
Jennings Plot)
4.8
Patient Result Report
The
quality control analysis report
(Levey Jen-
nings plot) shows the results of a selected reagent
over a period of time.
To generate a QA Analysis Report you have
to activate the QA-analysis by defining QA-
Labels in the Lot Specific Values while
preparing a worklist
(see chapter 4.3.4 on page 4-8)
.
A QA-Label is the label which is used in the assay
layout definition (e. g. NC, NC1, PC2 etc.).
1. Select File | New.
2. In the New dialogue box select the file type QA
Analysis Report.
3. In the Q.A. Analysis dialogue box select in
the box Controls the respective reagent.
4. Select in the box Batch Numbers the
respective bach number.
5. Select the period of time in the boxes Date
from and to.
6. Click OK to confirm the entries.
The system shows you the quality control analy-
sis report (Levey Jennings plot).
The patient result report shows a compact summary
of the patient results of all selected assays.
1. Select File | New.
2. In the New dialogue box, select the file type
Patient Result Report.
3. In the Patient Result dialogue box, select
All Patients or IDs between.
If you selected IDs between than enter the
first ID and the last ID in the two boxes.
4. Select up to five assays.
5. Select the period of time in the boxes Date
from and Date to.
Содержание Elisys Quattro
Страница 1: ...HUMAN ELISYS QUATTRO User Manual Elisys Quattro User Manual Cat No 16300 1...
Страница 6: ...TOC 4 ELISYS QUATTRO User Manual HUMAN...
Страница 26: ...2 12 ELISYS QUATTRO User Manual HUMAN System Basics...
Страница 66: ...5 8 ELISYS QUATTRO User Manual HUMAN Loading Unloading Disposal...
Страница 69: ...Service and Maintenance ELISYS QUATTRO User Manual HUMAN 6 3 Pipettor...
Страница 87: ...Technical Data ELISYS QUATTRO User Manual HUMAN 8 5...