background image

10

www.humancaregroup.com

Battery

Always ensure that battery is sufficiently charged 

before using this lift.  The battery will not charge if 

the emergency stop is engaged.

The battery in this device is located in the module 

along with the control box.

These batteries have a tamper proof seal that is 
located on the battery to make it easier to see if 
the battery has been tampered/replaced by a non-
authorized person. This seal should be replaced 
after battery replacement.

Changing the Battery

Remove all external cables 
from the control box.  Remove 
from frame.

On the back side of the 
module, unscrew and remove 
the 7 torx screws on the rear 
of the control box/battery 
module to open.

Disconnect the battery 
connectors and remove the 
old batteries.

Insert the new batteries in 
the same direction as the old 
batteries, with the connectors 
at the top and connect (Note: 
+= red, - = black).  Be careful 
not to bend the connectors.

 Do not cover the holes. 

 Ensure that none of the 

wires are squeezed or caught 
when reinstalling the cover. 

 Ensure the mains is not 

connected.

Ensure that the silicone string 
is in place and properly fitted 
before reinstalling the cover. 
Replace seal if necessary.

The silicone string separates the battery from the 
electronics and must not be damaged or modified 
as it will allow penetration of battery gas into the 
electronics components with risk of explosion.  Do 
not use if this seal is damaged in any way.

Charging the Battery

The battery should be charged minimum 8 hours 
before used for the first time.

Содержание Levanto

Страница 1: ...Levanto Part no 20000170 User Manual...

Страница 2: ...3 Product Benefits 3 Warnings 4 General Care and Use 5 Components 6 Technical Specifications 7 Product Label 8 Product Symbols 8 Understanding the Components 9 Usage Warnings 13 Usage Guidelines 14 Pr...

Страница 3: ...bols Safety Instructions WARNING symbol indicates a potentially hazardous situation which if not avoided could result in serious injury to the user or damage to property and or the device itself INFOR...

Страница 4: ...if Weight capacity is exceeded Product is damaged in any way The sling clips are not securely and adequately fastened You have not read and understood all warnings and instructions Warnings Do not tou...

Страница 5: ...ected by a professional with respect to the client s safety and needs Facility administrator and or risk management department should approve all patient handling protocols associated with this mobile...

Страница 6: ...humancaregroup com 4 1 2 3 5 7 6 Components 6 Front Wheel 7 Rear Wheel with brake 8 Hand Control not shown 9 Footplate not shown 1 Lift Arm 2 Push Handle 3 Battery Control Box Unit 4 Actuator 5 Shin...

Страница 7: ...e spreading legs IP Rating IPX4 Turning Radius 128cm 50 3 Wheels Front 10cm 4 Rear 12 5cm 5 with brake Battery 24V DC 2 x 12V 2 9Ah Li Ion accu 24V DC 2 9Ah Motor Lifting 24V DC 8A max 8kN 10 duty cyc...

Страница 8: ...35 EN 60601 1 Electrical Safety and EN 60601 1 2 EMC This product is CE marked Product Label Symbols on the unit and product label CE Marking Date of manufacture Manufactured for Human Care Maximum lo...

Страница 9: ...ox 1 Control functions up down 2 Emergency Stop 3 Mains socket Understanding the Hand Control Emergency Stop Use the emergency stop to turn off the lift when not in use to reduce the battery discharge...

Страница 10: ...ox battery module to open Disconnect the battery connectors and remove the old batteries Insert the new batteries in the same direction as the old batteries with the connectors at the top and connect...

Страница 11: ...th a very fast low noise and robust motor that is designed with a ball circulation spindle limit switches and automatic electric shut off in the event that the patient has come in contact with the arm...

Страница 12: ...rake is in the unlocked position line up the wheels with the chassis and the direction of travel Footplate The footplate is removable for hygiene control To remove lift footplate up to clean Clean wit...

Страница 13: ...Move furniture and other items that may impede the transfer in any way Explain the lifting process to the patient prior to lifting Ensure they have both the mental capability to understand and cooper...

Страница 14: ...cy button before use turn clockwise Usage Guidelines Moving the Patient Ensure that the patient is securely held in position during the entire move Do not move patient on an incline decline Unlock the...

Страница 15: ...roximately 2cm 0 8 below the patient s knees There should be no space between the leg and the pad Place the sling around the patient s back and pass the ends under the patient s arms Secure belt and t...

Страница 16: ...odium hypochlorite as recommended by the CDC 1 Weaker dilutions example 1 100 may be used but may not be in accordance with CDC recommendations Gloves and protective clothing should always be worn whe...

Страница 17: ...minimum It is recom mended that the following maintenance scheduled is maintained Checklist Prior to Each Use Battery is fully charged Sling shows no signs of wear Plastic components i e clips are not...

Страница 18: ...sible damage or wear All components are present eg battery There is no wear on the cables wires There is no damage to the chassis Operation The emergency stop is functioning The control box is functio...

Страница 19: ...ENGLISH www humancaregroup com 19...

Страница 20: ...il info ca humancaregroup com HUMAN CARE UNITED STATES 8006 Cameron Street Suite K Austin TX 78754 USA Phone 512 476 7199 Fax 512 476 7190 Email info us humancaregroup com HUMAN CARE AUSTRALIA Unit 4...

Отзывы: