Care instructions for the lift and accessories
Care instructions for the lift
•
The lift should be recharged every night
•
Wipe the ceiling lift using a slightly dampe
-
ned drying cloth. Washing up liquid can be
used
•
Never wash the ceiling lift using water or any
other fluids.
• NEVER use strong detergents on the ceiling
lift.
•
If the lift needs to be disinfected, DuPontTM
Virkon® can be used
•
The lift belt must be washed and disinfected
as instructed above
•
Activate the emergency stop if the ceiling lift
is not to be used for a few days. Otherwise
the batteries will run down
•
The lift must always be stored or moved
lying horizontally
Annual inspection
The lift must be inspected at least once a year
by someone with Human Care qualifications
and in accordance with Human Care’s directive.
Pay particular attention to parts that show wear.
The lift’s service indicator will flash yellow when
a service needs to be performed.
Daily inspection
•
Always check the lift belt and sling before
use. There must be no loose seams or other
signs of wear to lift belts, belts and fabric. If
there are, replace the sling or lift belt.
•
Check the ceiling lift for visible signs of
damage
•
Check that the lift is correctly mounted in
the distance strap or rail roller
Monthly inspection
Check the lift belts for damage or wear and tear.
Feed out enough of the lift belts to allow their
entire length to be checked. Contact a retailer
if the belt is damaged. The lift belt should be
replaced every second year.
Service
Installation, service and maintenance must only
be performed by personnel with Human Care
qualifications and in accordance with Human
Care’s directive. Only original spare parts may
be used.
Service agreements
Human Care invites you to sign a service agre-
ement for annual inspection and test loading.
Transport and storage
During transport, or when the lift is not going to
be used for some time, the mains switch must
be off or the emergency stop button pushed in,
otherwise the batteries will discharge. The lift
must be transported and stored lying horizon-
tally. The climatic conditions should be as fol
-
lows: ambient temperature from 0 °C to 40 °C,
relative ambient humidity from 30% to 80% and
ambient pressure from 790 hPa to 1060 hPa.
Product lifetime
The product´s expected lifetime is ten (10)
years or 10.000 lifts, if the product is used as
intended and maintained according to the
manufacturer´s instructions, depending on the
intensity of use and maximum load applied
during use. If the product label is no longer
legible, the product should be discarded.
Warranty and support
If you need information or support, please
contact your Human Care office,
www.humancaregroup.com, or distributor.