ENG

LIS

H

www.humancaregroup.com

11

Connect the power cord 
to the mains to charge 
the battery.  Make sure to 
disengage the emergency 
stop located at the side of the 
control box.

It is recommended that the batteries are charged 

at minimum once per week (depending on use).  

Regular charging prevents problems and extends the 

life of the battery.

Battery Lifespan

The number of lifting movements that can be made 

with a fully charged battery depends on the duration 

of the operation, the frequency of use, the patient 

weight and the age and condition of the battery.  

A battery that has not been used for a long period of 

time will slowly drain.  It is recommended that the 
battery is replaced every four years. 

These batteries 

have an anticipated lifespan of 2 to 4 years 
depending on usage.

Lifting Motor 

The Alzaro is equipped with a very fast, low-noise 

and robust motor that is designed with a ball 

circulation spindle, limit switches and automatic 

electric shut off in the event that the patient has 

come in contact with the arm or is sinking.

The motor is connected to the control box and 

allows the raising and lowering operation of the arm 

and the opening and closing operation of the legs.  

The lifting operations are performed with either the 

hand control (which is also connected to the control 

box) or from the control box itself.

Adjusting the legs

Always ensure there is sufficient space to spread the 

legs.  If the legs get stuck, this can cause damage to 

the lift or the surroundings.

To increase leg width, press 

on the open button on the 

hand control until desired 

spread is achieved.

To decrease leg width, press 

on the close button on the 

hand control until desired 

spread is achieved.

The Alzaro can be used with the legs in the wide 

or narrow position.  The choice between which 

position to use depends on the necessity and 

available space.

Содержание 11000490

Страница 1: ...Alzaro Part no 11000490 User Manual...

Страница 2: ...3 Product Benefits 3 Warnings 4 General Care and Use 5 Components 6 Technical Specifications 7 Product Label 8 Product Symbols 8 Understanding the Components 9 Usage Warnings 13 Usage Guidelines 14 Pr...

Страница 3: ...Safety Instructions WARNING symbol indicates a potentially hazardous situation which if not avoided could result in serious injury to the user or damage to property and or the device itself INFORMATIO...

Страница 4: ...s exceeded Product is damaged in any way The sling is not securely and adequately fastened You have not read and understood all warnings and instructions Warnings Do not touch any parts which may have...

Страница 5: ...ected by a professional with respect to the client s safety and needs Facility administrator and or risk management department should approve all patient handling protocols associated with this mobile...

Страница 6: ...ancaregroup com 4 1 2 3 6 5 7 Components 6 Front Wheel 7 Rear Wheel with brake 8 Hand Control not shown 9 Spreader Bar not shown 1 Lift Arm 2 Push Handle 3 Actuator 4 Battery Control Box Unit 5 Leg Ad...

Страница 7: ...reading legs IP Rating IPX4 Turning Radius 148cm 58 3 Wheels Front 10cm 4 Rear 12 5cm 5 with brake Battery 24V DC 2 x 12V 2 9Ah Li Ion accu 24V DC 2 9Ah Motor Lifting 24V DC 8 4A max 8kN 10 switch on...

Страница 8: ...nit and product label CE Marking Date of manufacture Manufactured for Human Care Maximum load Instructions Read the instructions and make sure you understand them before use Part number reference numb...

Страница 9: ...ox 1 Control functions up down 2 Emergency Stop 3 Mains socket Understanding the Hand Control Emergency Stop Use the emergency stop to turn off the lift when not in use to reduce the battery discharge...

Страница 10: ...battery module to open Disconnect the battery connectors and remove the old batteries Insert the new batteries in the same direction as the old batteries with the connectors at the top and connect Not...

Страница 11: ...th a very fast low noise and robust motor that is designed with a ball circulation spindle limit switches and automatic electric shut off in the event that the patient has come in contact with the arm...

Страница 12: ...e brake is in the unlocked position line up the wheels with the chassis and the direction of travel Manoeuvering the Alzaro Do not lift the Alzaro arm higher than is necessary to move it If you want t...

Страница 13: ...way Do not use if there are any signs of wear and damage on any part of the device Use product only on a level surface Do not use the lift on stairs escalators or an incline decline such as up or down...

Страница 14: ...control box will interrupt and stop the lifting process To activate press the red button De activate the emergency stop by turning clockwise Attaching the Spreader Bar The Alzaro is delivered with a t...

Страница 15: ...inue with lift If necessary advise patient that lift is commencing If using the self balancing crade attach sling clips to Alzaro lift arm clips Ensure all straps are straight and clips secure Using t...

Страница 16: ...odium hypochlorite as recommended by the CDC 1 Weaker dilutions example 1 100 may be used but may not be in accordance with CDC recommendations Gloves and protective clothing should always be worn whe...

Страница 17: ...ted Check the function on both the control box and the hand control if the control box works and the hand control does not then the hand control needs to be replaced The control box functions will not...

Страница 18: ...ccording to cleaning instructions Weekly Maintenance Check the operation of the lift and the operation of the emergency stop Check that all parts move easily and without any noise Charge the batteries...

Страница 19: ...visible damage or wear All components are present eg battery There is no wear on the cables wires There is no damage to the chassis Operation The emergency stop is functioning The control box is func...

Страница 20: ...l info ca humancaregroup com HUMAN CARE UNITED STATES 8006 Cameron Street Suite K Austin TX 78754 USA Phone 512 476 7199 Fax 512 476 7190 Email info us humancaregroup com HUMAN CARE AUSTRALIA Unit 4 4...

Отзывы: