Compliance (CDRH models only)
The CDRH model lasers (-1/300) are designed and manufactured to comply with the EC Low Voltage Directive and the
EC EMC Directive in the CDRH-compliant configuration of laser head, key control box, key and Cobolt-supplied power
supply. All equipment must be mounted on a common ground plane, such as an optical table. If any part of the
delivered equipment is replaced with a part not supplied by Cobolt or if the equipment is not properly grounded, the
system may not conform to CE / CDRH compliance standards listed here. Disabling any of the safety features nullifies
the CE marking and violates the laser safety standard.
The following harmonized and limits standards have been applied:
Electrical Safety:
EN 61010-1, IEC-61010-1, UL 61010-1 (Limited Energy System)
Laser Safety/Class
IEC-60825-1 , CDRH 21 CFR 1040.10 and 1040.11
EMC
4-2
IEC 61326-1
EN 55011
Electromagnetic Emission , Class B
FCC Part 15, subpart B, class B
Electromagnetic Immunity
–
Table 2 Requirements
EN 61000-4-2
Electrostatic Discharge
±4 kV contact discharge and
±2 kV, ±4 kV, ±8 kV air discharge
EN 61000-4-3
Radiated electromagnetic fields
80
–
1000 MHz 10 V/m with 80 % AM @ 1 kHz
1.4
–
2.7 GHz 10 V/m with 80 % AM @ 1 kHz
EN 61000-4-4
Fast transient / Burst
AC Power input port ±2,0 kV
EN 61000-4-5
Surge
AC Power input port ±0,5 kV, ±1,0 kV, ±2,0 kV Com. Mode
AC Power input port, ±0,5 kV, ±1,0 kV Diff. Mode
EN 61000-4-6
Conducted Immunity
10 V with 80 % AM @ 1 kHz
EN 61000-4-11 Dips and Interruptions
50 Hz and 60 Hz. Test voltages: 100 V and 230 V
RoHS
EU Directive 2011/65/EU
Contact your sales representative for a copy of the full Declaration of Conformity.