Setup mode
Chapter 4
OPERATION MANUAL
V2.3.0en/07.06.19//17.12
Unichiller®
55
Read the message and confirm it by tapping on “Yes”. The current control parameters will be overwritten.
Select the thermal fluid used from the list.
Confirm your choice by tapping on “OK”.
Read the message displayed and confirm it by tapping on “OK”.
Enter the filling volume using the number keypad that appears.
Confirm your entry by tapping on “OK”.
Tap twice on the “Arrow” touchbutton to return to the category “Temperature Control”.
Tap on the category “Process/Internal”.
Choose between the dialog entries “Internal” and “Process (Cascade)”.
Tap on “OK” to confirm your choice.
4.2.8.2
Sub-category: “Control Dynamics”
You can choose between a faster tuning performance with a possible and accepted small tempera-
ture overshoot and a tuning performance without temperature overshoot. The default setting is
“Fast, small overshoot”
.
The overshoot always refers to the leading temperature. Example: The process temperature, if activated, is the
leading temperature. In contrast, the bath or flow temperature must always be leading the process tempera-
ture. In order to achieve the best possible energy transfer, the greatest possible temperature difference be-
tween the bath or flow temperature and the process temperature is necessary (see figure "
»Representation of
an optimum process temperature control«
from page 51 in section
»Temperature control to process tempera-
). This always requires the maximum possible thermal fluid flow rate. In the setting
“Faster, small over-
shoot”
, the combination of high thermal fluid flow rate and the superbly designed control electronics hardly ever
results in an overshoot of the process temperature, while reaching the setpoint as quickly as possible. The oppo-
site to the mode
“Faster, small overshoot”
is available in the setting
“Without overshoot”.
The target tempera-
ture is approached more cautiously and thus aperiodic. It takes longer to adjust to the selected setpoint. The
statement “without overshooting” is only valid with little external disturbance influence. Please note the specifi-
cations on page 36 in section
»To be noted when planning the test«
.
The adjustment behavior can be selected at any time without reactivating the “Find control pa-
rameters”.
Internal, aperiodic
temperature control
Internal, dynamic
temperature control
with possible over-
shooting of the tem-
perature
Содержание Unichiller 400T
Страница 1: ...Unichiller...
Страница 2: ......
Страница 3: ...OPERATION MANUAL Unichiller...
Страница 4: ......
Страница 6: ...OPERATION MANUAL Unichiller V2 3 0en 07 06 19 17 12 6 Pilot ONE Layout of the Home screen...
Страница 85: ...Annex Chapter 9 OPERATION MANUAL V2 3 0en 07 06 19 17 12 Unichiller 85 9 Annex...