13
Figure
(
13
)
Wiring diagram (13) Inductance and reactance measurement wiring method, the
measurement line is inserted according to the color corresponding to the measurement
output end of the instrument, the yellow clip is clamped to one end of the busbar, the black
clip is clamped to the other end, and then the current measuring line is connected to the
instrument Tighten the interface, and the clamp sensor should be placed above the lead of
the reactor for measurement.
(7) Resistance measurement
By default, phase A is used for measurement, the power supply outputs yellow and black
binding posts, and the current input uses phase A clamps.
The measuring line is connected by the measuring output terminal of the instrument
according to the corresponding color. The red clip is clamped to one end of the resistor
and the black clamp is clamped to the other end. The clamp sensor should be placed on
the current line for measurement.
Note: The test wire is connected, the thick wire is connected to the current, and the thin
wire is connected to the voltage.
VIII.Precautions
1. Please read the instruction manual carefully before using this instrument, and check
that the wiring is correct and the grounding is good.