Huazheng HZRG-III Скачать руководство пользователя страница 16

13

Figure

13

Wiring diagram (13) Inductance and reactance measurement wiring method, the

measurement line is inserted according to the color corresponding to the measurement

output end of the instrument, the yellow clip is clamped to one end of the busbar, the black

clip is clamped to the other end, and then the current measuring line is connected to the

instrument Tighten the interface, and the clamp sensor should be placed above the lead of

the reactor for measurement.

(7) Resistance measurement

By default, phase A is used for measurement, the power supply outputs yellow and black

binding posts, and the current input uses phase A clamps.

The measuring line is connected by the measuring output terminal of the instrument

according to the corresponding color. The red clip is clamped to one end of the resistor

and the black clamp is clamped to the other end. The clamp sensor should be placed on

the current line for measurement.

Note: The test wire is connected, the thick wire is connected to the current, and the thin

wire is connected to the voltage.

VIII.Precautions

1. Please read the instruction manual carefully before using this instrument, and check

that the wiring is correct and the grounding is good.

Содержание HZRG-III

Страница 1: ...HZRG III Capacitance And Inductance Tester USER MANUAL Huazheng Electric Manufacturing Baoding Co Ltd...

Страница 2: ...ual and operate and maintain the instrument according to the manual to prolong its service life Just a light press the test will be completed automatically is the operating characteristics of this ins...

Страница 3: ...ntents I Overview 1 II Features 2 III Technical index 3 IV Conditions of Use 3 V Panel introduction 3 VI Instructions 4 VII Test wiring 10 VIII Precautions 13 IX After sales service 14 V Packing List...

Страница 4: ...itor bus and then measure with an old fashioned capacitor meter Since the capacitor bank is composed of dozens to hundreds of small capacitors it is necessary to disconnect the wires to measure the ca...

Страница 5: ...after the test is completed 2 During the measurement the instrument can display the measured voltage current power frequency impedance phase angle and other data while measuring the capacitance value...

Страница 6: ...k reading 3 0 1m 3 Dimensions 400mm 300mm 200mm IV Conditions of Use Environmental temperature 20 40 Environmental humidity 85 RH Power supply AC220V 10 50 1Hz V Panel introduction Figure 1 The four t...

Страница 7: ...connect and disconnect AC power USB U disk interface Instrument ground terminal RS232 extended spare VI Instructions When the instrument is connected to the test line and power cord as required turn o...

Страница 8: ...capacitance and inductance test If capacitance test is needed click the parameter setting icon button in the main menu screen to enter the capacitance measurement parameter setting screen as shown in...

Страница 9: ...nt the data click the print icon button to store the data click the save icon button if you still need to measure click the retest icon button click the exit icon button to return to the main interfac...

Страница 10: ...surement and wait for the instrument to automatically lock the screen In the locked screen state the user can save or print Click the corresponding icon button directly for each operation If you want...

Страница 11: ...t you can refer to the single phase capacitance measurement operation 4 Resistance test Connect the test wire and click the resistance test icon button on the main interface to directly perform resist...

Страница 12: ...ecommended that customers not change the setting data otherwise the test data will be inaccurate If you need to change it again it must be carried out under the guidance of our company s technicians a...

Страница 13: ...capacitive reactance of the capacitor under test the unit is ohm The inductive reactance of the reactor under test in ohm R the loss resistance of the tested product the unit is ohm D The loss factor...

Страница 14: ...t output terminal of the instrument according to the color corresponding to the A B C phases and then the current measurement line is tightened on the instrument interface and the clamp sensor should...

Страница 15: ...ee phase Yn shaped capacitance The three phase four wire Yn shaped measurement wiring is equivalent to the three phase Y type wiring or the single phase measurement method is adopted for the phase mea...

Страница 16: ...ult phase A is used for measurement the power supply outputs yellow and black binding posts and the current input uses phase A clamps The measuring line is connected by the measuring output terminal o...

Страница 17: ...contact with the tested product to avoid voltage open circuit 8 Do not remove the wiring before the test is completed during the measurement process 9 If there are problems that cannot be solved pleas...

Отзывы: