2
The temperature of the heating furnace is high in the process of measuring this
instrument, and it needs to be attended by someone.
3.Technical parameter
Applicable standards: gb/t 3536-2008, gb/t 267-88 and other standards
Working power supply: AC220V 10% 50Hz
Heating power: 600W (standby)
Overall power: maximum 750W
Ambient temperature: 5-40 ° C
Overall dimension: 380x360x360mm
Weight of main engine: About kg
Range: room temperature ~ 400
℃
Resolution: 0.1
℃
Repeatability: ≤ 4
℃
4.Instrument characteristics
Display interface: 8-inch color LCD touch screen.
Temperature measuring element: PT100 (L-type armor).
Ignition mode: the diameter of electronic ignition ball is 3-4mm (the temperature of the
fireball can be adjusted).
Flash point detection: UV flame sensor
Output mode: embedded micro thermal printer.
Safety protection: when the sample is on fire, the fire cover automatically covers the
cup to extinguish the fire, and can also conduct the sample test of ignition.
Wind protection: polymethylacrylic protective cover.
The PID closed-loop temperature control technology is used to realize the good
control of the temperature rising speed.
Strong air cooling device is adopted to quickly cool the heating furnace.
100 groups of experimental data can be stored.
II Instrument structure and components
The overall structure of the instrument is as follows: