11
Lifting mechanism: control the lifting or stopping of the lifting mechanism
Fire sweeping mechanism: control the fire sweeping mechanism to sweep or stop
Flash point detection: click the "flash point test stop" switch to convert the switch text
into "flash point test start", and then close the fire source to the experimental cup, and
the instrument will give a beep.Click the "flash point test start" switch to convert the
switch text to "flash point test stop", and the flash point detection will be stopped, but it
will not affect the flash point detection in the experiment process
Ignition mechanism: control the preheating, ignition or off of ignition mechanism, note
that ignition power must be turned on
Hot bath control: control the power of the heating rod, start the heating, and display the
current sample temperature and the hot bath temperature
Cooling fan: controls the start or stop of the hot bath cooling fan.
Ignition power supply: control the power supply of ignition mechanism on or off.
Exit: cancel flash point detection, cooling fan control, ignition power control, ignition
mechanism control in the current interface, and return to the main interface.
6.Instrument correction
The instrument corrects the temperature and the air pressure correction is shown in the
following figure: