5
●
Test setup interface
SetThe test time
Battery Indicator
Set Short Current
Set the filter
Button function tip
Soft key
Description
OFF
Power Off (No operation, 3 minutes automatic shutdown)
H.V
、
1000V
、
2500V
、
5000V
TIME
SetThe test time: 01:00>10:00>30:00
Switching software keys to a new function, or from other function
to return
I
Set Short Current
,
1 mA step selection Usually according to the
maximum, can also according to the regulations stipulated by the
short circuit current value choice.
Filter
Set the filter:
0
:
without filtering; (no interference case recommended)
1
:
hardware low-pass filter; (applicable to most situations)
2
:
software Low pass filter;High resistance (greater than 100G
recommended choose)
3
:
ha software low-pass filter;
(high resistance is greater than 1T recommend choose)
HELP
About Tester message
Battery Indicator
:
、 、 、 、
When the battery voltage shows just one more case should be timely charge, when
the battery displaying flashing is automatically cut off the power at any time, should
be recharged.
charge
:
AC socket provided meet the requirements of the AC power supply for panel,
you can recharge equipment charging, the power indicator, display the battery label,
in turn, from
(0) to (full) showed that when filled, displays (full).
Contrast Settings interface: LCD display gray level change due to temperature
changes, may affect the electronic contrast adjustment using the instrument Settings