background image

5

Test setup interface

Set test voltage

SetThe test time

Battery Indicator

Set Short Current

Set the filter

Button function tip

Soft key

Description

OFF

Power Off (No operation, 3 minutes automatic shutdown)

H.V

Set test voltage: 500V

1000V

2500V

5000V

TIME

SetThe test time: 01:00>10:00>30:00
Switching software keys to a new function, or from other function
to return

I

Set Short Current

1 mA step selection Usually according to the

maximum, can also according to the regulations stipulated by the
short circuit current value choice.

Filter

Set the filter:
0

without filtering; (no interference case recommended)

1

hardware low-pass filter; (applicable to most situations)

2

software Low pass filter;High resistance (greater than 100G

recommended choose)
3

ha software low-pass filter;

(high resistance is greater than 1T recommend choose)

HELP

About Tester message

Battery Indicator

、 、 、 、

When the battery voltage shows just one more case should be timely charge, when

the battery displaying flashing is automatically cut off the power at any time, should

be recharged.

charge

AC socket provided meet the requirements of the AC power supply for panel,

you can recharge equipment charging, the power indicator, display the battery label,

in turn, from

(0) to (full) showed that when filled, displays (full).

Contrast Settings interface: LCD display gray level change due to temperature

changes, may affect the electronic contrast adjustment using the instrument Settings

Содержание HZJY-5K-I

Страница 1: ...HZJY 5K I Insulation Resistance Tester Huazheng Electric Manufacturing Baoding Co Ltd...

Страница 2: ...l and operate and maintain the instrument according to the manual to prolong its service life Just a light press the test will be completed automatically is the operating characteristics of this instr...

Страница 3: ...Contents I Safe Use Items 1 II Feature 2 III Specification Parameter 3 IV Introduction Of Instrument 4 V Instrument Using 7 VI Packing List 8...

Страница 4: ...tions If you don t comply measurement may cause personal injury and equipment damage Dangerous for improper operation can lead to serious or fatal damage Before use please wear insulated gloves Please...

Страница 5: ...ltage insulation resistance tester with 5 range 500V 1000V 2500V 5000V test up to 5T Strict accordance with the safety standards for design Insulation resistance range 5T 5 kv Short circuit current ca...

Страница 6: ...our size L x W x H 26cm x 20cm x 16cm weight 3kg Test voltage accuracy the nominal value 100 to 110 Voltage measurement accuracy 5 3V Current test range 10mA current measure precision 5 0 2nA short ci...

Страница 7: ...g 2 AC ON 3 Test Port E G L 4 LCD display 5 Key Key and function Key numbered from left to right the first for a total of five off down the power Key key number Description 1 Press the switch on butto...

Страница 8: ...pass filter applicable to most situations 2 software Low pass filter High resistance greater than 100G recommended choose 3 hardware software low pass filter high resistance is greater than 1T recomm...

Страница 9: ...OK Confirm to enter the current menu function or to determine the current Settings and exit Help menu system HELP Record the serial number of product production date version information Insulation tes...

Страница 10: ...e polarization index after 600 seconds Polarization index calculation method PI R600 R60 test connection points for attention 1 Tested to confirm product safety grounding try not charged 2 Confirm the...

Страница 11: ...8 VI Packing List No Item Qty 1 Main engine 1 2 Power line 1 3 Red test line 1 4 Black test line 1 5 Green test line 1...

Отзывы: