Safe Use Items

Design and manufacture of the instrument and testing achieves IEC61010 safety

standards (electronic measuring product safety requirements), this manual including the

safety of the instrument use and ensure the safety of equipment state, users must follow

the warnings and safety regulations Please read the following instructions before use.

Warning

Instrument output voltage, before use, read and understand the instructions of

operational guidelines.

Please take manual save you to check at any time.

Must be instructed to use the instrument.

Understand and comply with the safety operating instructions.

Must strictly abide by the above instructions.

If you don't comply, measurement may cause personal injury and equipment damage.

Dangerous (for improper operation can lead to serious or fatal damage)

Before use, please wear insulated gloves.

Please do not over AC/DC600V measurement circuit.

Please do not test in flammable places, spark could cause an explosion.

please do not hand wet or damp operators in instrument surface operation.

Please don't charged the connection test line low.

Low in the measurement, or do not touch the circuit being measured immediately after the

test, may result in electric shock accidents.

Low test line or port found easy to damage the insulation properties of dirt or carbon

please stop test.

Warning (for improper operation exists the possibility of a severe or fatal damage)

If the instrument is abnormal, please stop using it. For example: instrument damage or
bare metal parts.
Please do not install in the device replacement parts or any unauthorized modification,
maintenance, please contact with me.
Make sure all wires and instrument testing port connection is firm.

Note (for improper operation may cause personal injury or equipment damage)

Содержание HZJY-10K-IA

Страница 1: ...HZJY 10K IA Insulation Resistance Tester User Manual Huazheng Electric Manufacturing Baoding Co Ltd...

Страница 2: ...al and operate and maintain the instrument according to the manual to prolong its service life Just a light press the test will be completed automatically is the operating characteristics of this inst...

Страница 3: ...Contents I Safe Use Items 1 II Features 2 III Technical specifications 3 IV Instrument introduction 4 V Instrument use 5 VI Appendix 7...

Страница 4: ...erious or fatal damage Before use please wear insulated gloves Please do not over AC DC600V measurement circuit Please do not test in flammable places spark could cause an explosion please do not hand...

Страница 5: ...nt and capacitance as well as dielectric discharge test DD step voltage test SV ramp test the instrument automatically boosts Until the set value it is qualified If it is not reached there is a breakd...

Страница 6: ...l value Output voltage monitoring accuracy 5 10V Voltage measurement range AC 30 600V 50HZ 60HZ DC 30 600V Voltage measurement accuracy 2 3dgt Current test range 10mA Current measurement accuracy 5 0...

Страница 7: ...utton 6 High voltage indicator Keys and function description There are 5 buttons in total from left to right The first one is the power on button in the shutdown state Button No Instruction 1 Power on...

Страница 8: ...When charging the power indicator light is on and the battery indicator on the display screen displays from 0 grid to full grid When it is full it displays full grid System help menu description Help...

Страница 9: ...stance values of 15S 60S and 600S and calculates the polarization index after 600 seconds Polarization index calculation method PI R600 R60 DD test This test is suitable for the diagnosis of multilaye...

Страница 10: ...should be connected to the G terminal of the meter with a test wire to eliminate the leakage current caused by the surface of the test product The measurement error ensures the accuracy of the test Es...