VI. Precautions
1. The selection, placing and electrode distance of the oil sample before testing, shall be
accord with national standards and industry standards.
2. The operators or other personnel are strictly forbidden to touch the casing after the
power is switched on, to avoid accidents.
3. The power shall be cut off immediately if any abnormal event is found during the
operation.
VII. Maintenance
1. This equipment shall not be exposed in moist environment.
2. Keep the oil cup and the electrodes clean. Fill the cup with fresh transformer oil for
protection during its idle. Check the electrode distance and check the tightness between
the electrode tip and electrode bar screw thread before the cup is used again.
VIII. Oil Cup Cleaning Method And Common Fault
Clearances
1
、
Oil Cup Cleaning Method
⑴
Wipe the electrode surfaces and bars again and again with clean silk cloth.
⑵
Adjust the electrode distance with standard gauge.