8
work
Stir motor is
too noisy
Stir paddle and stationary
shaft are not homocentric,
friction occurs.
Open exterior case of instrument,
loosen the screw fixing motor, adjust
position of motor support board to
make stir shaft and stationary shaft
homocentric
Fire go out
when ignition
1
.
Test flame too small
2
.
Disturbing air flow
1
.
Adjust flame size
2
.
Prevent air flow
V.Maintenance
5.1
The instrument should be placed at a draft and dry place.
5.2
After test, the test cup and all parts related to the sample steam should be clean
5.3
The instrument should be covered by cloth after being used.
5.4
The warranty is one year after delivery.