8
In Figure 5, when the cursor is in the data query, press the select key to scroll up, and
press the start key to scroll down. Move the cursor to print, press the select key according
to the prompts to print the currently displayed data, move the cursor to the U disk for
storage, and press the select key according to the prompts to save the current displayed
data to the USB.
Ⅵ
. Wiring method
Note: All connecting surfaces of the test clamp should be in reliable contact with the
sample. If oxidation is found on the contact surface, the oxide layer on the contact surface
should be removed first.
1. Wiring method: As shown in Figure 6, connect the dedicated test wire to the
corresponding I+ and I- terminals according to the color red to red, black to black, and
tighten the thin power wire into V+, In the V- socket, two clamps clamp the two ends of
the tested product.
2. Measurement: After completing the wiring in Figure 6, turn on the power switch and
select the test time for testing. The instrument will automatically complete charging,
steady current and resistance measurements.
Ⅶ
.Safety measures
1. The operator should have common sense in the use of general electrical equipment or
instruments.
2. After the test, press the reset button to turn off the power switch. Note: It is strictly
forbidden to disassemble and assemble the test cables when powered.