Troubleshooting
5.7. Enable/disable of the highest and lowest
temperature cursor
As shown in the figure, after selecting “ ” (cold hotspot), press the
“ ” button in the navigation key to enter the cold hotspot setting.
If you encounter with any problem when using the thermal image device,
please use the following table for repair.If the problems are not solved,
please cut off the power supply and contact with the manufacturer.
Press “ ” /“ ” key to select “enable” or “disable” the option.
Then press “ SELECT/ENTER ” key to determine selection.
After the setting is completed,press “ ” key to return. Press
“ / MENU” key to exit from the menu.
Spot
Off
On
24
h
Failure phenomenon
The thermal imaging
device cannot start
The thermal imaging
device powers off
automatically
Failure cause
The battery is not
installed
The power of the battery
is used up
Solution
Install the battery
Replace with new
battery or charge it
The power of the
battery is used up
Replace with new
battery or charge it
The time set for
automatic power off
is due
Restart or change the
time for automatic power
off after restarting
(refer to 5.1)