1. SPECIFICATIONS

1.1 General Specifications
Display:                         3 ½ digit liquid crystal display (LCD) 20mm 
                                      height with a maximum reading of 1999.
Over range Indication : (1) or (-1) is displayed at MSD 
Low Battery :                Displayed when the battery voltage drops is low. 
Safety standard :          Meets the requirement of IEC 1010 for installation
                                      category II, 1000V phase to earth.  
Measurement rate:      2.5 measurements per second.
Operating condition :  0°C to + 50°C, 0-75% RH
Storage conditions  :   -20°C to + 60°C, 0-80% RH with battery removed.
Accuracy  :                    ± (3% rdg + 2dgt ) Accuracy specifications at 23± 
                                     5°C, less than 75% RH.
Power Supply :             9 volts (6 x 1.5V) battery, Eveready type 1015 or 
                                     equivalent. OR Power Adaptor 9V, 1.5 A
Dimensions:                 139 x 99 x 54mm (approx)
Weight :                       395gms including battery, Eveready type 1015 or 
                                     equivalent.
Accessories :                Test leads (pair) X 1, 1.5V battery (installed) X 
                                     carrying case X 1.
Dielectric Strength :     3.5kV @ 50Hz for 1 min. between input terminals 
                                     and case.

Insulation Resistance : More than 50MΩ at 500V between circuit and case.

                                   1.2 Electrical Specifications

2. SAFETY INFORMATION

  

1. The circuit under test must be de energized and isolated before 
    connections are made.
2. Circuit connections must not be touched during a test.
3. After insulation tests capacitive circuits must be allowed to be 
    discharged.
4. Test leads must be in good order, clean and have no cracked or 
    broken insulation.
5. Do not push the test button before all connections and preparations 
    are done. The instrument must only be used by suitable trained and 
competent persons.   

Warning and safety symbols     Caution  refer to this 
manual before using the meter.      Dangerous 
voltages. C€ Comply with IEC1010 – 1. 

Digital Insulation Tester

Model: DIT-90

1

Accuracies are ±(% reading plus number of digits) at 23 ± 5°C, and 
humidity of less than 75% RH

Model NO.

1000V

DIT-90A

DIT-90B

DIT-90C

DIT-90D

DIT-90E

1000V

500V

500V

2000MΩ

200MΩ

2000MΩ 200MΩ

1MΩ

0.1MΩ

1 MΩ

0.1 MΩ 

100V

200MΩ
0.1 MΩ 

Rated Voltage
Range

MAX.

Resolution
Max. No. load

Voltage

± 25% of rated voltage

<2 mA

± (3%rdg + 2dgt)

Short Circuit

Current

Accuracy