![HT VEGA74 Скачать руководство пользователя страница 35](http://html1.mh-extra.com/html/ht/vega74/vega74_user-manual_2178126035.webp)
VEGA74
EN - 34
10. TECHNICAL SPECIFICATIONS
Accuracy is calculated as: ±[%r (no. of digits) * resolution] at 23°C, <80%RH.
10.1. TECHNICAL CHARACTERISTICS AUX AND LEAKAGE SECTIONS
Leakage current (input I1
– STD clamp)
FS AC Clamp [A]
Resolution [A]
Accuracy
1
0.1mA
(1
rdg + 20dgt)
1< FS < 10
0.01A
10
FS < 100
0.1A
100
FS
1000
1A
Environmental parameters
Measurement
Range
Resolution
Accuracy
°C
-20.0
60.0°C
0.1°C
(2%rdg + 2digits)
°F
-4.0
140.0°F
0.1°F
RH%
0.0%
100.0%RH
0.1%RH
DC voltage
0.1mV
1.0V
0.1mV
Lux
0.001
20.00lux (*)
0.001
0.02Lux
0.1
2.0klux (*)
0.1
2Lux
1
20.0klux (*)
1
20Lux
(*) Accuracy of the luxmetric probe according to Class AA