VEGA
76
EN - 45
13.1.6. Measurement of harmonics
Range Accuracy Resolution
DC
÷
25h
±
(5.0%+2digit)
26h
÷
33h
±
(10.0%+2digit)
34h
÷
49h
±
(15.0%+2digit)
0.1V / 0.1A
13.1.7. Frequency measurement
Instrument set to 50Hz
Range Accuracy Resolution
47
÷
53
±
(1.0%+1digit)
0.1Hz
Instrument set to 60Hz
Range Accuracy Resolution
57
÷
63.6
±
(1.0%+1digit)
0.1Hz
13.1.8. Compliance
VEGA 76 complies with the Standards prescribed for:
•
class 2 EN61036 – static counters of active energy
•
class 3 IEC1268 – static counters of reactive energy
13.1.9. Temperature drift
Temperature drift:
0,1 x accuracy/K
13.1.10. Safety
The instrument complies with: EN 61010-1 + A2(1996)
Insulation:
Class 2
Pollution: 2
Overvoltage category:
CAT III 300V~; CAT II 350V~ (Phase-Earth)
CAT III 600V~ (Phase-Phase)
13.1.11. General
Features
Mechanical characteristics
Dimensions:
225(L) x 165(W) x 105(H) mm
Weight (including battery):
1.5kg
Power supply
Battery type:
6 batteries x 5V series AA LR6
Indication of low battery:
the symbol "" is displayed when the battery is low
Battery life:
approx. 50 hours
External power supply:
Use only the provided power supplier
Display
Specifications:
dot matrix with backlight
Resolution
128 x 128 dots (16384 dots)
Dot size
0.5mm x 0.5mm
Visible area
73mm x 73mm
Sampling
Sampling speed:
156.25usec @ 50Hz.
No. of samples per period:
128
Содержание VEGA 76
Страница 1: ...Copyright HT ITALIA 2004 Release EN 2 00 08 06 2004 VEGA 76 User s Manual...
Страница 2: ......