INSTRUCTION FOR USE
HSM 600
INSTRUMENT TABLE
Warning!
Before using this product, read the instructions carefully. This contains important information re-garding user safety.
Страница 1: ...INSTRUCTION FOR USE HSM 600 INSTRUMENT TABLE Warning Before using this product read the instructions carefully This contains important information re garding user safety ...
Страница 2: ...ce This device is intended for use in professional healthcare facilities such as medical practices and op tometrist and optician premises It must not be operated in the vicinity of high frequency surgical sys tems and within shielded rooms of ME systems for nuclear spin tomography Portable radio frequency devices such as mobile phones or other RF telephone accessories including antennas may affect...
Страница 3: ...version 13 7 6 Installation of optional trays PC tray or standard tray 13 7 7 Screw sets 14 8 Commissioning Operation Decommissioning 15 8 1 Commissioning HSM 600 15 8 2 Operation HSM 600 15 8 3 Transport of the HSM 600 16 8 3 1 Short distances 16 8 3 2 Longer distances 16 8 4 Decommissioning HSM 600 16 8 5 Technical data 17 8 6 Dimensions Weights 17 9 Maintenance fuse replacement cleaning wipe di...
Страница 4: ...10 Figure 7 Mounting the Base Plate preparation 11 Figure 8 Mounting the base plate inserting and tightening the 4 Screws 11 Figure 9 Mounting the base plate wiring and securing with cable ties 12 Figure 10 Leveling HSM 600 Instrument table 12 Figure 11 Assembly Octopus 600 instrument table special version 13 Figure 12 Mounting optional shelves 13 Figure 13 Screws and cable ties 14 Figure 14 Fuse ...
Страница 5: ...pening Warning Always follow the instructions Electrical appliances of protection class I protective earth CE Certificate of Conformity Manufacturer Year Serial number Reference Application part of type B Note Disposal Warning Electricity Forbidden In case of non observance danger Important notes Table 1 Symbols ...
Страница 6: ...o means that only suitable and tested hoists may be used Uneven ground Note that on an uneven floor or similar The goods must be secured professionally against slipping sideways Transport on EURO pallet The table components are usually delivered on EURO pallets Only suitable pallet trucks or heavy duty wheeled boards may be used to transport or move the pallets be used During transport and storage...
Страница 7: ...sons The power cable must only be connected to a correctly installed power supply This must be installed in accordance with the VDE guidelines Risk of electric shock In order to avoid the risk of electric shock this device may only be connected to a supply network with protective conductor The instrument table must be checked according to EN 62353 IEC 62353 before commissioning In the event of dam...
Страница 8: ...20610 120V 7220611 230V LAN 7220612 120V LAN SL PS LED SMALL 7220601 230V 7220602 120V 7220603 230V LAN 7220604 120V LAN SL RM 02 SMALL 7220605 230V 7220606 120V 7220607 230V LAN 7220608 120V LAN OCTOPUS 600 7220625 230V 7220626 120V 7220627 230V LAN 7220628 120V LAN OCTOPUS 900 7220621 230V 7220622 120V 7220623 230V LAN 7220624 120V LAN LENSTAR 900 7220617 230V 7220618 120V 7220619 230V LAN 72206...
Страница 9: ...ed in 230 V or 120 V with or without LAN cabling Please compare the power voltage with the existing voltage on the label Columns with LAN option Columns with LAN option have an integrated LAN cable prefabricated with RJ45 socket in the column below and RJ45 extender on the column above LAN cables must be preassembled according to 8connecting the lifting column earthing cable Place the HSM 600 tabl...
Страница 10: ...cable lugs of the device column see arrow Figure 2 Connecting lifting column connect earthing cable secure cable with cable tie Then secure the cable with a cable tie Figure 3 Connecting the lifting column mounting the cable whip and the earthing cable Insert the cable and the ground cable into the cable duct Figure 4 Connecting the lifting column mounting the metal cover 2 x M6x16 ...
Страница 11: ...ocker switch to raise the lifting column Operation HSM 600 page15 Figure 5 Connection column connection power cable and mounting column Place the lifting column on the 8 predrilled holes 4 holes each on the right and left side Now insert the 8 screws 4x M6x16 on the front panel 4x M6x25 incl washer at the rear of the column Note Fix the lifting column with the 8 screws Tighten the screws firmly Fi...
Страница 12: ...persons are required for safe installation Figure 7 Mounting the Base Plate preparation Before further installation please be within easy reach 4 screws M6x45 incl washer and hexagon socket screwdriver Figure 8 Mounting the base plate inserting and tightening the 4 Screws Inserting the 4 screws M6x45 incl washers Note Tighten all 4 screws firmly and check that they are properly tightened 4x M6x45 ...
Страница 13: ...asher are preassembled 7 4 Leveling HSM 600 Instrument table The HSM 600 tables are factory leveled and are supplied with the named washers and mounted wheels on the base plates The mounted wheels are secured with screw locking lac quer If the HSM 600 table plate is not in the scale e g uneven floor the se can be levelled using 6 washers Ø 8 4 mm A maximum of 3 washers per wheel are permitted Note...
Страница 14: ...l trays PC tray or standard tray Figure 12 Mounting optional shelves Note A suitable set of screws is supplied with each shelf Insert the green yellow earthing cable into the cable duct and plug the earthing cable into one of the two cable lugs on the lifting column Then tie the cable with a cable tie Insert the 4 screws M4x10 and tighten them firmly For double installation from PC storage and sta...
Страница 15: ... and cable ties Base bag 2 x cable ties 1x SCREW SET 1 1x SCREW SET 2 SCREW SET 1 base plate 4x screws M6x45 incl 4 washers DIN 912 SCREW SET 2 lifting column 4x screws M6x16 DIN 912 for control panel on device column 4x screws M6x25 incl washers DIN 912 for lifting column ...
Страница 16: ...peration WARNING Risk to clamp Before changing the table height make sure that the movement path is free of extremities of the patient e g legs or of obstacles If this is not done there is a risk of clamping the patient s thighs or other extremities or other things 8 2 Operation HSM 600 The HSM 600 instrument table has an electromotive lifting column The control panel is mounted at the top of the ...
Страница 17: ...ap the brakes back into place Release the brakes on the 2 wheels and move the table to the desired position 8 3 2 Longer distances Watch your step Note cable length Limited cable lengths can damage the cable or plug by pulling it After transport to the desired position engage the brakes again Place the table board in the lowest position and disconnect the power cable from the power supply Loosen t...
Страница 18: ...HSM 600 for Slit Lamp small Tabletop dimension mm 860 x 440 Height min max in mm 680 980 Weight kg 46 Type HSM 600 for Slit Lamp large Tabletop dimension mm 1150 x 440 Height min max in mm 680 980 Weight kg 49 Type HSM 600 for BX 900 Tabletop dimension mm 1150 x 440 Height min max in mm 680 980 Weight kg 49 Type HSM 600 for Lenstar Tabletop dimension mm 1150 x 440 Height min max in mm 680 980 Weig...
Страница 19: ...ve the country specific regulations 9 2 Fuse replacement Fuse replacement Only fuses as described in the corresponding table under chapter Technical data may be used The fuses F1 and F2 are located on the right side of the control panel below the tabletop Page 17 Voltage 230 V 2 x T4 0 AL 250V Voltage 120 V 2 x T6 3 AL 250V Disconnect the power cord Unlock the fuse holder with a screwdriver and pu...
Страница 20: ...tix com Bacillol 30 Foam Ref 981307 Bacillol 30 Tissues Ref 981312 Manufacturer BODE Chemie GmbH D 22525 Hamburg www bode chemie com 10 Legal regulations classification disposal standards 10 1 Legal regulations The HSM 600 instrument tables are designed and manufactured according to the following standards EN 60601 1 EN 60601 1 2 Production inspection assembly maintenance and repair in accordance ...
Страница 21: ...ument tables are constructed in such a way that the generation and emission of electromagnetic disturbances is limited to such an extent that other devices are not disturbed in their intended operation and themselves exhibit adequate immunity to electromagnetic disturbances 11 2 Interfering emissions Standard table 1 Table 6 Interference emissions Standard table 1 12 Classification Medical Product...
Страница 22: ...e doc 21 22 11 06 2019 HS DOMS GmbH Langenschader Straße 67 D 07318 Saalfeld Germany Phone 49 3671 54 50 0 Fax 49 3671 54 50 20 info hs doms eu www hs doms com Subject to changes in design and form without prior notice ...