![HP VXI 75000 C Series Скачать руководство пользователя страница 104](http://html.mh-extra.com/html/hp/vxi-75000-c-series/vxi-75000-c-series_users-and-scpi-programming-manual_157876104.webp)
104
Programming the HP E1415 for PID Control
correct built-in table will be used for reference junction compensation.
Reference junction compensation is based on the reference junction
temperature at the time the custom channel is measured. For more
information see “Thermocouple Reference Temperature Compensation” on
page 68.
Custom Reference
Temperature EU
Conversions
The HP E1415 can measure reference junction temperatures using custom
characterized RTDs and thermistors. The custom EU table generated for the
individually characterized transducer is loaded to the appropriate channel(s)
using the DIAG:CUST:PIEC command. Since the EU conversion from this
custom EU table is to be considered the "reference junction temperature",
the channel is linked to this EU table using the command
[SENSe:]FUNCtion:CUSTom:REFerence [<range>,](@<ch_list>).
This command uses the custom EU conversion to generate the reference
junction temperature as explained in the section “Thermocouple Reference
Temperature Compensation” on page 68.
Creating Conversion
Tables
Contact your Hewlett-Packard System Engineer for more information on
Custum Engineering Unit Conversion for your application.
Loading Custom EU
Tables
There is a specific location in the E1415’s memory for each channel’s EU
Conversion table. When standard EU conversions are specified, the E1415
loads these locations with EU conversion tables copied from its non-volatile
FLASH Memory. For Custom EU conversions you must load these table
values using either of two SCPI commands.
Loading Tables for Linear Conversions
The DIAGnostic:CUSTom:LINear <table_range>,<table_block>,
(@<ch_list>) command downloads a custom linear Engineering Unit
Conversion table to the HP E1415 for each channel specified.
•
<table_block> is a block of 8 bytes that define 4, 16-bit values. SCPI
requires that <table_block> include the definite length block data
header. C-SCPI adds the header for you.
•
<table_range> specifies the range of input voltage that the table
covers (from -<table_range> to +<table_range>). The value you
specify must be within 5% of: .015625 | .03125 | .0625 | .125 | .25 | .5 |
1 | 2 | 4 | 8 | 16 | 32 | 64.
•
<ch_list> specifies which channels will have this custom EU table
loaded.
Usage Example
Your program puts table constants into array
table_block
DIAG:CUST:PIEC
table_block
,1,(@132:163)
send table for chs 32-63 to
HP E1415
SENS:FUNC:CUST:PIEC 1,1,(@132:163)
link custom EU with chs 32-63
Содержание VXI 75000 C Series
Страница 2: ......
Страница 16: ...16 ...
Страница 18: ......
Страница 30: ...30 Getting Started Chapter 1 Notes ...
Страница 32: ...32 Field Wiring Chapter 2 Figure 2 1 Channel Numbers at SCP Positions ...
Страница 44: ...44 Field Wiring Chapter 2 Figure 2 11 HP E1415 Terminal Module ...
Страница 54: ...54 Field Wiring Chapter 2 Notes ...
Страница 61: ...Programming the HP E1415 for PID Control 61 Chapter 3 Programming Overview Diagram ...
Страница 124: ...124 Creating and Running Custom Algorithms Chapter 4 Figure 4 2 Algorithm Operating Sequence Diagram ...
Страница 136: ...136 Creating and Running Custom Algorithms Chapter 4 Notes ...
Страница 152: ...152 Algorithm Language Reference Chapter 5 Notes ...
Страница 304: ...304 HP E1415 Command Reference Chapter 6 Command Quick Reference Notes ...
Страница 308: ...308 Specifications Appendix A Thermocouple Type E 200 800C SCPs HP E1501 02 03 ...
Страница 309: ...Specifications 309 Appendix A Thermocouple Type E 200 800C SCPs HP E1508 09 ...
Страница 310: ...310 Specifications Appendix A Thermocouple Type E 0 800C SCPs HP E1501 02 03 ...
Страница 311: ...Specifications 311 Appendix A Thermocouple Type E 0 800C SCPs HP E1509 09 ...
Страница 312: ...312 Specifications Appendix A Thermocouple Type E Extended SCPs HP E1501 02 03 ...
Страница 313: ...Specifications 313 Appendix A Thermocouple Type E Extended SCPs HP E1508 09 ...
Страница 314: ...314 Specifications Appendix A Thermocouple Type J SCPs HP E1501 02 03 ...
Страница 315: ...Specifications 315 Appendix A Thermocouple Type J SCPs HP E1508 09 ...
Страница 316: ...316 Specifications Appendix A Thermocouple Type K SCPs HP E1501 02 03 ...
Страница 317: ...Specifications 317 Appendix A Thermocouple Type R SCPs HP E1501 02 03 ...
Страница 318: ...318 Specifications Appendix A Thermocouple Type R SCPs HP E1508 09 ...
Страница 319: ...Specifications 319 Appendix A Thermocouple Type S SCPs HP E1501 02 03 ...
Страница 320: ...320 Specifications Appendix A Thermocouple Type S SCPs HP E1508 09 ...
Страница 321: ...Specifications 321 Appendix A Thermocouple Type T SCPs HP E1501 02 03 ...
Страница 322: ...322 Specifications Appendix A Thermocouple Type T SCPs HP E1508 09 ...
Страница 323: ...Specifications 323 Appendix A 5K Thermistor Reference SCPs HP E1501 02 03 ...
Страница 324: ...324 Specifications Appendix A 5K Thermistor Reference SCPs HP E1508 09 ...
Страница 325: ...Specifications 325 Appendix A RTD Reference SCPs HP E1501 02 03 ...
Страница 326: ...326 Specifications Appendix A RTD SCPs HP E1501 02 03 ...
Страница 327: ...Specifications 327 Appendix A RTD SCPs HP E1508 09 ...
Страница 328: ...328 Specifications Appendix A 2250 Thermistor SCPs HP E1501 02 03 ...
Страница 329: ...Specifications 329 Appendix A 2250 Thermistor SCPs HP E1508 09 ...
Страница 330: ...330 Specifications Appendix A 5K Thermistor SCPs HP E1501 02 03 ...
Страница 331: ...Specifications 331 Appendix A 5K Thermistor SCPs HP E1508 09 ...
Страница 332: ...332 Specifications Appendix A 10K Thermistor SCPs HP E1501 02 03 ...
Страница 333: ...Specifications 333 Appendix A 10K Thermistor SCPs HP E1508 09 ...
Страница 334: ...334 Specifications Appendix A Notes ...
Страница 346: ...346 Glossary Appendix C Notes ...
Страница 388: ...388 Generating User Defined Functions Appendix F Notes ...
Страница 438: ...438 Index Writing the algorithm 129 values to CVT elements 120 values to the FIFO 121 Z ZERO CALibration ZERO 194 ...