ENWW
Calibrating the spectrophotometer 33
Calibrating the spectrophotometer
Performing a full calibration
Before using the spectrophotometer for the first time, and once a month thereafter,
perform a full calibration as follows:
1
From the
Utilities
menu, click
Calibrate Spectrophotometer
, and
Full Calibration
.
2
Follow the instructions in the series of message windows that appear, clicking
OK
after each message window.
Note
For the Black Calibration test, make sure that you measure the dark area until the four vibrating
measurements end completely. This takes approximately 15 seconds.
If the
Black calibration successful
and the
White calibration
successful
messages appear on the screen, the full calibration was successful. If
not, repeat this procedure one more time before calling the customer engineer.
Performing the daily calibration
Calibrate the spectrophotometer once a day.
1
From the
Utilities
menu, click
Calibrate Spectrophotometer
, and
Frequent
Calibration
.
2
Follow the instructions in the window that appears.
Note
For the White Calibration test, make sure that you measure the white reference area until the
single vibrating measurement ends completely. This takes approximately two seconds.
If the
Calibration successful
message appears on the screen, the daily
calibration was successful. If not, repeat this procedure one more time before calling
the customer engineer.
Содержание IndiChrome
Страница 1: ...HP IndiChrome Ink Mixing System User guide ...
Страница 2: ...HP IndiChrome Ink Mixing System User guide ...
Страница 7: ...6 Contents ENWW ...
Страница 63: ...62 Chapter 4 General procedures ENWW ...
Страница 90: ...ENWW 89 7Help This chapter contains the following sections z Online help z Useful links z Program updates ...
Страница 92: ...ENWW Program updates 91 ...
Страница 93: ...92 Chapter 7 Help ENWW ...
Страница 103: ...EN 102 ...