HP HP 8901A Скачать руководство пользователя страница 1

H

HP 8901A Modulation Analyzer

150 kHz to 1300 MHz

Product Overview

Outstanding signal
characterization

Mobile Radio

The HP 8901A Modulation Analyzer combines all the capabilities
necessary for making accurate transmitter measurements. It
can be used in making all the tests listed below. It can also be
used to test modules and subassemblies from either the trans-
mitter or receiver. For reliability, input power protection reduces
the chance of accidentally damaging the modulation analyzer by
connecting it directly to a high power transmitter.

• Carrier power

• Carrier frequency and stability

• AM depth

• FM deviation

• Hum and noise

• Incidental AM or FM

• Modulation limiting
   - Instantaneous
   - Steady state

• Audio frequency response

Maintenance and Metrology

Accurately measuring modulation has long been a problem for
metrology laboratories. The HP 8901A Modulation Analyzer
helps solve this problem two ways. First, it provides an extremely
accurate method of measuring AM depth and FM deviation, and
it recovers the modulation with little degradation for making
measurements such as modulation distortion. Second, the optional
calibrators provide a precise modulation source for calibration.
There are many metrology and maintenance uses for the modula-
tion analyzer. They include:

• Signal generator calibration

• Modulation calibration standard

• VCO and VCXO characterization

Applications

Frequency

Range: 

150 kHz to 1300 MHz

Resolution:

10 Hz below 1000 MHz,
100 Hz above 1000 MHz

Input Level:

Automatic Mode: 

20 dBm to

+30 dBm

Manual Mode:

 

Typically

60 dBm to +30 dBm

Power

Display: 

Peak envelope power

Range:

 1 milliwatt to 1 watt

Accuracy:

 

Typically

 

±1.5 dB

Input Power Protection:

>25 watts

Modulation

Frequency Modulation

Rates: 

20 Hz to 200 kHz

Deviations:

 to 400 kHz

Accuracy: 

±1% of reading ±1 digit

for rates 30 Hz to 100 kHz

Amplitude Modulation

Rates:

 20 Hz to 100 kHz

Depths:

 to 99%

Accuracy: 

±1% of reading ±1 digit

for rates 50 Hz to 50 kHz and
depths >5%

Phase Modulation

Rates:

 200 Hz to 20 kHz

Deviations:

 to 400 radians

Accuracy:

 ±3% of reading ±1 digit

Measurements

*

*All parameters describe performance in automatic operation or properly set manual conditions.  Specifications

describe the instrument’s warranted performance.  Supplemental characteristics (shown in italics) are intended to

provide information useful in applying the instrument by giving typical, but non-warranted, performance parameters.

Содержание HP 8901A

Страница 1: ...distortion Second the optional calibrators provide a precise modulation source for calibration There are many metrology and maintenance uses for the modula tion analyzer They include Signal generator calibration Modulation calibration standard VCO and VCXO characterization Applications Frequency Range 150 kHz to 1300 MHz Resolution 10 Hz below 1000 MHz 100 Hz above 1000 MHz Input Level Automatic M...

Страница 2: ...tic system to make all these measurements under remote control Single Key Measurements The HP 8901A Modulation Analyzer features easy to use controls In automatic operation all major func tions are selected by pushing a single key No manual tuning or range selection is needed The modu lation analyzer automatically tunes to the input signal adjusts for proper signal level selects the appropriate me...

Страница 3: ... 1000 MHz Sensi tivity is 25 dBm 12 mVrms below 650 MHz and 20 dBm 22 mVrms above 650 MHz Besides normal frequency measurement capabilities the analyzer s counter has several additional capabilities Like most frequency counters it will mea sure signals over a wide dynamic range 50 dB 22 mVrms to 7 Vrms and is protected from damage for signals up to 35 Vrms Unlike many frequency counters however it...

Страница 4: ...tec tion to prevent damage from the accidental application of exces sive power This is a common cause of damage in equipment such as the modulation analyzer used to measure transmitters The modulation analyzer is tested for inputs up to 25 watts Protec tion is provided by limiting diodes and an RF relay When excessive power is applied the relay opens and protects sensitive compo nents and the anal...

Страница 5: ...o be accurately measured The modulation analyzer contains four de emphasis networks that can be used in addition to the audio filters These are the ones commonly used in FM communi cations 25 50 75 and 750 µs When selected the de emphasis networks always affect the de modulated output The user can select whether or not the de em phasis network affects the devia tion measured and indicated by the d...

Страница 6: ...pecial functions can also be used to set any measurement range or instrument function They can be used to select either of two inter nal IF s the one normally used for frequencies above 10 MHz or a narrow IF where rates and devia tions are more restricted but selectivity is increased All instru ment functions not set using these special functions remain in auto matic mode This allows the user to s...

Страница 7: ...ions Besides improving the accuracy of these measure ments it greatly reduces the time involved in making measure ments Also the optional calibra tors provide a new level of modulation standard accuracy and help ensure accurate mea surements They also enhance the calibration of the modulation analyzer itself Because the modulation analyzer is useful for characterizing all types of signals in gener...

Страница 8: ...nty period Hewlett Packard Company will at its option either repair or replace products which prove to be defective For warranty service or repair this product must be returned to a service facility designated by HP Buyer shall prepay shipping charges to HP and HP shall pay shipping charges duties and taxes for products returned to HP from another country HP warrants that its software and firmware...