prog.book : chapter1.fm 69 Mon Feb 15 09:24:37 1999
Power Meter Remote Operation
Getting the Best Speed Performance
HP E4418B/E4419B Programming Guide
1-69
•
the hardware platform being used.
•
the programming environment being used.
Note that it is unlikely that 200 readings/sec can be achieved when:
•
you are using a 700 series HPUX workstation.
•
you are using a low end PC.
•
you are using a graphical programming environment (such as HP
VEE).
Dual Channel Considerations
With the dual channel instrument, consideration must be taken of what
operation is required on both channels. Both channels can achieve 20
readings/sec simultaneously, and 40 readings/sec simultaneously, but 200
readings/sec is not achievable on both channels at the same time. If only
single channel measurements are required, then the other channel should
be set to standby mode and not triggered.
The throughput for a channel set in the 200 readings/sec mode will be
affected by the speed mode of the other channel. However, in a situation
where fast measurements are required on one channel and slow
measurements on the other, it will be possible to perform more than one
measurement cycle on the fast channel for every measurement on the slow
channel. For example, if channel A is set to 40 readings /sec and channel B
is set to 20 readings/sec, it is possible to construct a loop with 2 reads from
channel A and one from channel B and still achieve the set readings per
second.
Содержание E4418B
Страница 22: ...prog book progLOF doc 12 Mon Feb 15 09 24 37 1999 Contents 12 HP E4418B E4419B Programming Guide ...
Страница 24: ...prog book progLOT doc 14 Mon Feb 15 09 24 37 1999 Contents 14 HP E4418B E4419B Programming Guide ...
Страница 25: ...prog book chapter1 fm 1 Mon Feb 15 09 24 37 1999 1 Power Meter Remote Operation ...
Страница 138: ...prog book ch2meas fm 1 Mon Feb 15 09 24 37 1999 2 MEASurement Instructions ...
Страница 200: ...prog book ch3calc fm 1 Mon Feb 15 09 24 37 1999 3 CALCulate Subsystem ...
Страница 228: ...prog book ch4cal fm 1 Mon Feb 15 09 24 37 1999 4 CALibration Subsystem ...
Страница 242: ...prog book ch5disp fm 1 Mon Feb 15 09 24 37 1999 5 DISPlay Subsystem ...
Страница 260: ...prog book ch6form fm 1 Mon Feb 15 09 24 37 1999 6 FORMat Subsystem ...
Страница 264: ...prog book ch7mem fm 1 Mon Feb 15 09 24 37 1999 7 MEMory Subsystem ...
Страница 294: ...prog book ch8out fm 1 Mon Feb 15 09 24 37 1999 8 OUTput Subsystem ...
Страница 302: ...prog book ch9sens fm 1 Mon Feb 15 09 24 37 1999 9 SENSe Subsystem ...
Страница 360: ...prog book ch10stat fm 1 Mon Feb 15 09 24 37 1999 10 STATus Subsystem ...
Страница 384: ...prog book ch11syst fm 1 Mon Feb 15 09 24 37 1999 11 SYSTem Subsystem ...
Страница 420: ...prog book ch12trig fm 1 Mon Feb 15 09 24 37 1999 12 TRIGger Subsystem ...
Страница 434: ...prog book ch13unit fm 1 Mon Feb 15 09 24 37 1999 13 UNIT Subsystem ...
Страница 442: ...prog book ch14serv fm 1 Mon Feb 15 09 24 37 1999 14 SERVice Subsystem ...
Страница 452: ...prog book ch15ieee fm 1 Mon Feb 15 09 24 37 1999 15 IEEE488 2 Command Reference ...
Страница 481: ...prog book progIX doc 6 Mon Feb 15 09 24 37 1999 Index 6 HP E4418B E4419B Programming Guide ...