Power Meter Remote Operation
Making Measurements
HP E4418B/E4419B Programming Guide
1-15
Using MEASure?
The simplest way to program the power meter for measurements is by
using the
MEASure?
query. However, this command does not offer much
flexibility. When you execute the command, the power meter selects the
best settings for the requested configuration and immediately performs
the measurement. You cannot change any settings (other than the
expected power value, resolution and with the HP E4419B the
measurement type) before the measurement is taken. This means you
cannot fine tune the measurement, for example, you cannot change the
filter length. To make more flexible and accurate measurements use the
CONFIGure
command. The measurement results are sent to the output
buffer.
MEASure?
is a compound command which is equivalent to an
ABORT
, followed by a
CONFigure
, followed by a
READ?
.
MEASure? Examples
The following commands show a few examples of how to use the
MEASure?
query to make a measurement. It is advisable to read through
these examples in order as they become increasingly more detailed. These
examples configure the power meter for a measurement (as described in
each individual example), automatically place the power meter in the
“wait-for-trigger” state, internally trigger the power meter to take one
reading, and then sends the reading to the output buffer.
These examples give an overview of the
MEASure?
query. For further
information on the
MEASure?
commands refer to the section
“MEASure[1|2] Commands” starting on page 2-49 .
Example 1 - The Simplest Method
The following commands show the simplest method of making single
channel (for example A or B) measurements. Using
MEAS1?
will result in
an upper window measurement, and
MEAS2?
in a lower window
measurement. The channel associated with the window can be set using
the source list parameter (see example 2), or will default as in this
example (See “HP E4419B only” on page 18.).
MEAS1?
specifies window
MEAS2?
Содержание E4418B
Страница 22: ...prog book progLOF doc 12 Mon Feb 15 09 24 37 1999 Contents 12 HP E4418B E4419B Programming Guide ...
Страница 24: ...prog book progLOT doc 14 Mon Feb 15 09 24 37 1999 Contents 14 HP E4418B E4419B Programming Guide ...
Страница 25: ...prog book chapter1 fm 1 Mon Feb 15 09 24 37 1999 1 Power Meter Remote Operation ...
Страница 138: ...prog book ch2meas fm 1 Mon Feb 15 09 24 37 1999 2 MEASurement Instructions ...
Страница 200: ...prog book ch3calc fm 1 Mon Feb 15 09 24 37 1999 3 CALCulate Subsystem ...
Страница 228: ...prog book ch4cal fm 1 Mon Feb 15 09 24 37 1999 4 CALibration Subsystem ...
Страница 242: ...prog book ch5disp fm 1 Mon Feb 15 09 24 37 1999 5 DISPlay Subsystem ...
Страница 260: ...prog book ch6form fm 1 Mon Feb 15 09 24 37 1999 6 FORMat Subsystem ...
Страница 264: ...prog book ch7mem fm 1 Mon Feb 15 09 24 37 1999 7 MEMory Subsystem ...
Страница 294: ...prog book ch8out fm 1 Mon Feb 15 09 24 37 1999 8 OUTput Subsystem ...
Страница 302: ...prog book ch9sens fm 1 Mon Feb 15 09 24 37 1999 9 SENSe Subsystem ...
Страница 360: ...prog book ch10stat fm 1 Mon Feb 15 09 24 37 1999 10 STATus Subsystem ...
Страница 384: ...prog book ch11syst fm 1 Mon Feb 15 09 24 37 1999 11 SYSTem Subsystem ...
Страница 420: ...prog book ch12trig fm 1 Mon Feb 15 09 24 37 1999 12 TRIGger Subsystem ...
Страница 434: ...prog book ch13unit fm 1 Mon Feb 15 09 24 37 1999 13 UNIT Subsystem ...
Страница 442: ...prog book ch14serv fm 1 Mon Feb 15 09 24 37 1999 14 SERVice Subsystem ...
Страница 452: ...prog book ch15ieee fm 1 Mon Feb 15 09 24 37 1999 15 IEEE488 2 Command Reference ...
Страница 481: ...prog book progIX doc 6 Mon Feb 15 09 24 37 1999 Index 6 HP E4418B E4419B Programming Guide ...