Chapter 3
3-19
Programming Examples
Performing Internal Self-alignment
Performing Internal Self-alignment
This C programming example does the following:
• Open an HP-IB session at address 18.
• Clear the analyzer.
• Reset the analyzer.
• VISA function sets the time out to infinite for this specified session.
• Initiate a self-alignment.
• Query for operation complete.
• Query for results of self-alignment.
• Report the results of the self-alignment.
• Close the session.
Example:
#include <stdio.h>
#include <stdlib.h>
#include <ctype.h>
#include <string.h>
#include "visa.h"
#define hpESA_IDN_E4401B
"Hewlett-Packard, E4401B"
#define hpESA_IDN_E4411B
"Hewlett-Packard, E4411B"
Visession
defaultRM, viESA;
ViStatus
errStatus;
ViChar
cIdBuff[256]={0};
char
cEnter = 0;
int
iResult = 0;
/*Set the input port to the 50 MHz amplitude reference*/
void Route50MHzSignal()
{
viQueryf(viESA, "*IDN?\n", "%t", &cIdBuff);
iResult = (strncmp( cIdBuff, hpESA_IDN_E4401B,
strlen(hpESA_IDN_E4401B)) && strncmp( cIdBuff,
hpESA_IDN_E4411B, strlen(hpESA_IDN_E4411B)));
if( iResult == 0 )
{
Содержание E4401B
Страница 6: ...vi ...
Страница 14: ...xiv Contents ...
Страница 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Страница 65: ...3 1 3 Programming Examples ...
Страница 118: ...3 54 Chapter3 Programming Examples Entering Amplitude Correction Data Close the session viClose viESA viClose defaultRM ...
Страница 151: ...4 1 4 Programming Command Cross References ...
Страница 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Страница 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Страница 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Страница 368: ...7 1 7 Error Messages ...
Страница 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...