![HP E4401B Скачать руководство пользователя страница 52](http://html.mh-extra.com/html/hp/e4401b/e4401b_programmers-manual_161548052.webp)
2-22
Chapter 2
HP ESA Status Registers
Using the Analyzer Status Registers
Questionable Status Power
event enable register
The Questionable Status Power Condition Register contains the
following bits:
Figure 2-9
Bit
Description
0
Reserved. These bits are not used by the analyzer, but are for
future use with other HP products.
1
A 1 in this bit position indicates that the source (tracking
generator) output is unlevelled.
2
A 1 in this bit position indicates that the local oscillator (LO) in
the source (tracking generator) is unlevelled.
3
A 1 in this bit position indicates that the analyzer local oscillator
(LO) is unlevelled.
4
A 1 in this bit position indicates that the 50 MHz amplitude
reference signal is unlevelled.
5
Reserved. These bits are not used by the analyzer, but are for
future use with other HP products.
6
A 1 in this bit position indicates that the input overload protection
(HP ESA models E4401B and E4411B only) is tripped.
7
A 1 in this bit position indicates that the first local osillator (LO)
output is unlevelled.
8–14
Unused. These bits are always set to 0.
15
Always Zero (0).
Bit Number
7
8
9
10
11
12
13
14
15
6 5 4 3 2 1 0
Reserved
50
MHz
Osc
Unleveled
LO
Unleveled
Source
LO
Unleveled
Source
Unleveled
Reserved
Unused Unused Unused
Unused
Unused
LO
Out
Unleveled
Unused
Unused
Input
Overload
Tripped
Description
Always
Zero
(0)
ck770a
Questionable
Power Condition Register
Status
STATus:QUEStionable:POWer:CONDition?
Содержание E4401B
Страница 6: ...vi ...
Страница 14: ...xiv Contents ...
Страница 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Страница 65: ...3 1 3 Programming Examples ...
Страница 118: ...3 54 Chapter3 Programming Examples Entering Amplitude Correction Data Close the session viClose viESA viClose defaultRM ...
Страница 151: ...4 1 4 Programming Command Cross References ...
Страница 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Страница 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Страница 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Страница 368: ...7 1 7 Error Messages ...
Страница 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...