7-10
Chapter 7
Error Messages
Error Message Types
Error Message Types
Events do not generate more than one type of error. For example, an
event that generates a query error will not generate a device-specific,
execution, or command error.
These errors indicate that the instrument output queue control has
detected a problem with the message exchange protocol described in
IEEE 488.2, Chapter 6. Errors in this class set the query error bit (bit
2) in the event status register (IEEE 488.2, section 11.5.1). These errors
correspond to message exchange protocol errors described in IEEE
488.2, 6.5. In this case:
• Either an attempt is being made to read data from the output queue
when no output is either present or pending, or
• data in the output queue has been lost.
These errors indicate that the instrument parser detected an IEEE
488.2 syntax error. Errors in this class set the command error bit (bit 5)
in the event status register (IEEE 488.2, section 11.5.1). In this case:
• Either an IEEE 488.2 syntax error has been detected by the parser
(a control-to-device message was received that is in violation of the
IEEE 488.2 standard. Possible violations include a data element
which violates device listening formats or whose type is
unacceptable to the device.), or
• an unrecognized header was received. These include incorrect
device-specific headers and incorrect or unimplemented IEEE 488.2
common commands.
201 to 799:
Device-Specific
Errors
These errors indicate that a device operation did not properly complete,
possibly due to an abnormal hardware or firmware condition. These
codes are also used for self-test response errors. Errors in this class set
the device-specific error bit (bit 3) in the event status register (IEEE
488.2, section 11.5.1).
The <error_message> string for a positive error is not defined by SCPI.
A positive error indicates that the instrument detected an error within
the HP-IB system, within the instrument firmware or hardware, during
the transfer of block data, or during calibration.
Содержание E4401B
Страница 6: ...vi ...
Страница 14: ...xiv Contents ...
Страница 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Страница 65: ...3 1 3 Programming Examples ...
Страница 118: ...3 54 Chapter3 Programming Examples Entering Amplitude Correction Data Close the session viClose viESA viClose defaultRM ...
Страница 151: ...4 1 4 Programming Command Cross References ...
Страница 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Страница 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Страница 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Страница 368: ...7 1 7 Error Messages ...
Страница 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...