![HP E4401B Скачать руководство пользователя страница 277](http://html.mh-extra.com/html/hp/e4401b/e4401b_programmers-manual_161548277.webp)
Chapter 5
5-123
Language Reference
SYSTem Subsystem
Save User Preset
:SYSTem:PRESet[:USER]:SAVE
Saves the current instrument conditions as the user preset condition.
Front Panel
Access:
System, Power On/Preset, Save User Preset
Speaker Control
:SYSTem:SPEaker[:STATe] OFF|ON|0|1
:SYSTem:SPEaker[:STATe]?
Turns the internal speaker on or off.
Factory Preset
and *RST:
Off
Front Panel
Access:
Det/Demod, Speaker On Off
Set Time
:SYSTem:TIME <hour>,<minute>,<second>
:SYSTem:TIME?
Sets the time of the real-time clock of the instrument.
Hour must be an integer 0 to 23.
Minute must be an integer 0 to 59.
Second must be an integer 0 to 59.
Front Panel
Access:
System, Time/Date, Set Time
SCPI Version Query
:SYSTem:VERSion?
Returns the SCPI version number with which the instrument complies.
Содержание E4401B
Страница 6: ...vi ...
Страница 14: ...xiv Contents ...
Страница 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Страница 65: ...3 1 3 Programming Examples ...
Страница 118: ...3 54 Chapter3 Programming Examples Entering Amplitude Correction Data Close the session viClose viESA viClose defaultRM ...
Страница 151: ...4 1 4 Programming Command Cross References ...
Страница 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Страница 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Страница 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Страница 368: ...7 1 7 Error Messages ...
Страница 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...