![HP E4401B Скачать руководство пользователя страница 240](http://html.mh-extra.com/html/hp/e4401b/e4401b_programmers-manual_161548240.webp)
5-86
Chapter 5
Language Reference
SENSe Subsystem
SENSe:FREQuency Subsection
Center Frequency
[:SENSe]:FREQuency:CENTer <freq>
[:SENSe]:FREQuency:CENTer?
Set the center frequency.
Factory Preset
and *RST:
HP ESA E4401B, E4411B: 750 MHz
HP ESA E4402B, E4403B: 1.5 GHz
HP ESA E4404B: 3.35 GHz
HP ESA E4405B: 6.6 GHz
HP ESA E4407B, E4408B: 13.25 GHz
Range:
HP ESA E4401B, E4411B: –80 MHz to 1.58 GHz
HP ESA E4402B, E4403B: –80 MHz to 3.10 GHz
HP ESA E4404B: –80 MHz to 6.78 GHz
HP ESA E4405B: –80 MHz to 13.3 GHz
HP ESA E4407B, E4408B: –80 MHz to 27.0 GHz
Default Unit:
Hz
Front Panel
Access:
FREQUENCY/Channel, Center Freq
Center Frequency Step Size Automatic
[:SENSe]:FREQuency:CENTer:STEP:AUTO OFF|ON|0|1
[:SENSe]:FREQuency:CENTer:STEP:AUTO?
Specifies whether the step size is set automatically based on the span.
Factory Preset
and *RST:
On
Front Panel
Access:
FREQUENCY/Channel, CF Step Auto Man
Center Frequency Step Size
[:SENSe]:FREQuency:CENTer:STEP[:INCRement] <freq>
[:SENSe]:FREQuency:CENTer:STEP[:INCRement]?
Specifies the center frequency step size.
Factory Preset
and *RST:
Span/10
Содержание E4401B
Страница 6: ...vi ...
Страница 14: ...xiv Contents ...
Страница 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Страница 65: ...3 1 3 Programming Examples ...
Страница 118: ...3 54 Chapter3 Programming Examples Entering Amplitude Correction Data Close the session viClose viESA viClose defaultRM ...
Страница 151: ...4 1 4 Programming Command Cross References ...
Страница 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Страница 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Страница 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Страница 368: ...7 1 7 Error Messages ...
Страница 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...