Chapter 5
5-53
Language Reference
INITiate Subsystem
INITiate Subsystem
The INITiate subsystem is used to control the initiation of the trigger.
Refer to the TRIGger and ABORt subsystems for related commands.
Continuous or Single Measurements
:INITiate:CONTinuous OFF|ON|0|1
:INITiate:CONTinuous?
Selects whether the trigger system is continuously initiated or not.
This command affects sweep if not in a measurement, and affects
trigger when in a measurement. A “measurement” refers to any of the
functions under the
MEASURE
key. This corresponds to continuous
sweep or single sweep operation when not in a measurement, and
continuous measurement or single measurement operation when in a
measurement.
NOTE
When not in a measurement, this command does the following:
• When ON at the completion of each sweep cycle, the sweep system
immediately initiates another sweep cycle.
• When OFF, the sweep system remains in an “idle” state until
CONTinuous is set to ON or an
:INITiate[:IMMediate]
command is received. On receiving the
:INITiate[:IMMediate]
command, it will go through a single sweep cycle, and then return to
the “idle” state.
• The query returns 1 or 0 into the output buffer. 1 is returned when
there is continuous sweeping. 0 is returned when there is only a
single sweep.
When in a measurement, this command does the following:
• When ON at the completion of each trigger cycle, the trigger system
immediately initiates another trigger cycle.
• When OFF, the trigger system remains in an “idle” state until
CONTinuous is set to ON or an
:INITiate[:IMMediate]
command is received. On receiving the
:INITiate[:IMMediate]
command, it will go through a single trigger cycle, and then return to
the “idle” state.
• The query returns 1 or 0 into the output buffer. 1 is returned when
there is continuous triggering. 0 is returned when there is only a
single trigger.
Factory Preset: Continuous
Содержание E4401B
Страница 6: ...vi ...
Страница 14: ...xiv Contents ...
Страница 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Страница 65: ...3 1 3 Programming Examples ...
Страница 118: ...3 54 Chapter3 Programming Examples Entering Amplitude Correction Data Close the session viClose viESA viClose defaultRM ...
Страница 151: ...4 1 4 Programming Command Cross References ...
Страница 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Страница 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Страница 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Страница 368: ...7 1 7 Error Messages ...
Страница 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...