Chapter 5
5-13
Language Reference
CALCulate Subsystem
limit line is not affected by this command.
Front Panel
Access:
Display, Limits, Properties, Limits Fixed Rel
Set Limit Line X-axis Units
:CALCulate:LLINe:CONTrol:DOMain FREQuency|TIME
:CALCulate:LLINe:CONTrol:DOMain?
Selects how the limit line segments are defined: according to frequency,
or according to the sweep time setting of the spectrum analyzer.
NOTE
Changing this setting deletes all existing limit data from the analyzer.
In other words, if a limit line has already been defined, changing the
type clears the existing limit line.
Factory Preset
and *RST:
not affected by Preset
Remarks:
For TIME, the limit line segments are placed on the
spectrum analyzer display with respect to the sweep
time setting of the analyzer, with 0 at the left edge of
the display.
For FREQuency, segments are placed according to the
frequency that is specified for each segment.
Front Panel
Access:
Display, Limits, Properties, X Axis Units Freq Time
Control Limit Line Frequency Interpolation
:CALCulate:LLINe[1]|2:CONTrol:INTerpolate:TYPE
LOGarithmic|LINear
:CALCulate:LLINe[1]|2:CONTrol:INTerpolate:TYPE?
Selects the type of interpolation done for the frequency values of the
designated limit line when comparing to measured data. This only
applies in the frequency mode. This function does not work in zero span
(for example, in the time domain.)
Default Unit:
Log
Remarks:
Once this function is defined, the selected type is
persistent. Persistent means that it retains the setting
previously selected, even through a power cycle.
Front Panel
Access:
Display, Limits, Modify, Freq Interp Log Lin
Содержание E4401B
Страница 6: ...vi ...
Страница 14: ...xiv Contents ...
Страница 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Страница 65: ...3 1 3 Programming Examples ...
Страница 118: ...3 54 Chapter3 Programming Examples Entering Amplitude Correction Data Close the session viClose viESA viClose defaultRM ...
Страница 151: ...4 1 4 Programming Command Cross References ...
Страница 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Страница 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Страница 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Страница 368: ...7 1 7 Error Messages ...
Страница 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...