3-40
Chapter 3
Programming Examples
Creating a Limit Line
Creating a Limit Line
This C programming example does the following:
• Open an HP-IB session at address 18.
• Clear the analyzer.
• Reset the analyzer.
• Set the input port to the 50 MHz amplitude reference.
• Delete any current limit line.
• Define the upper limit line to have frequency/amplitude pairs.
• Define the lower limit line to have frequency/amplitude pairs.
• Turn the limit line test function on.
• Set the analyzer to a center frequency of 50 MHz, span to 20 MHz,
and resolution bandwidth to 1 MHz.
• Turn the limit line test function on.
• Set the analyzer reference level to 0 dBm.
• Activate the 50 MHz Alignment signal.
• Check to see if limit line passes or fails. It should pass
• Pause for 5 seconds.
• Deactivate the 50 MHz alignment signal
• The limit line test should fail.
• Close the session.
Example:
#include <stdio.h>
#include <stdlib.h>
#include <math.h>
#include <conio.h>
#include <ctype.h>
#include <string.h>
#include <windows.h>
#include "visa.h"
#define YIELD Sleep(5000)
Содержание E4401B
Страница 6: ...vi ...
Страница 14: ...xiv Contents ...
Страница 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Страница 65: ...3 1 3 Programming Examples ...
Страница 118: ...3 54 Chapter3 Programming Examples Entering Amplitude Correction Data Close the session viClose viESA viClose defaultRM ...
Страница 151: ...4 1 4 Programming Command Cross References ...
Страница 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Страница 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Страница 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Страница 368: ...7 1 7 Error Messages ...
Страница 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...