:ENABle?
[:EVENt]?
STATus:OPERation:CONDition?
STATus:OPERation:CONDition?
returns the contents of the Operation Status
Condition Register. Reading the register does not affect its contents.
Parameters
Returns: 16-bit signed integer (int16).
Comments
•
Bit 0 of this register (decimal 1) is the Calibrating bit. This bit is set when:
–
the
CALibration:CONFigure:RESistance?
,
CALibration
n
:CONFigure:VOLTage?
,
or
CALibration
n
:CONFigure:CURRent?
command has been received and,
–
measurements with the corresponding
CALibration:VALue:RESistance
,
CALibration
n
:VALue:VOLTage?
, or
CALibration
n
:VALue:CURRent?
commands
are not complete.
•
*RST Condition: clears all conditions.
•
*CLS
and
STATus:PRESet
do not affect bits in the Status Operation Condition
Register.
STATus:OPERation:ENABle
STATus:OPERation:ENABle <
mask
>
specifies which bits of the Operational
Status Event Register are OR’d together to create the Operational Status bit (bit 7)
in the Summary Register of the Status Byte.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
<
mask
>
numeric (int16) or
non-decimal numeric
-32768 to +32767 or #H0000 to #HFFFF
none
Comments
•
Only bit 0 (decimal 0) is used or defined in the Operation Register. Bit 0 is
defined as the Calibrating bit. A 1 in this bit indicates the module is currently in
the calibration mode.
•
The non-decimal numeric forms are the #H, #Q, or #B formats specified by
IEEE-488.2.
•
*RST
clears all conditions.
•
*CLS
and
STATus:PRESet
do not affect bits in the Operation Status Condition
Register.
STATus
95 HP E1418A SCPI Command Reference
Chapter 3
Содержание E1418A
Страница 6: ...Notes 6 Contents HP E1418A 8 16 Channel D A Converter Module ...
Страница 10: ...Notes 10 HP E1418A User s Manual ...
Страница 12: ...12 HP E1418A User s Manual ...
Страница 105: ...TRIGger 105 HP E1418A SCPI Command Reference Chapter 3 ...
Страница 111: ...Notes HP E1418A Command Quick Reference 111 HP E1418A SCPI Command Reference Chapter 3 ...
Страница 135: ...135 HP E1418A Register Based Programming Appendix B ...
Страница 157: ...Notes 156 HP E1418A Error Messages Appendix C ...
Страница 164: ...Figure D 4 Resistance Calibration Connections CAL Appendix D Voltage Current Output Adjustment 163 ...
Страница 170: ...Notes Appendix D Voltage Current Output Adjustment 169 ...
Страница 174: ...Figure E 1 8 Channel Disassembly 172 Configuration and Disassembly Appendix E ...
Страница 175: ...Figure E 2 16 Channel Disassembly Appendix E Configuration and Disassembly 173 ...
Страница 177: ...Figure E 3 Plug on Channel Locations Figure E 4 Installing Plug on Modules Appendix E Configuration and Disassembly 175 ...
Страница 192: ...Notes 192 HP E1418A 8 16 Channel D A Converter Module Index ...