CALibration
n
:CONFigure:CURRent?
CALibration
n
:CONFigure:CURRent?
configures the channel for current mode
calibration and returns a
0
or
1
to indicate the channel’s readiness for calibration.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
CALibration
n
keyword
1 to 8 or 1 to 16
1
Returns: an unsigned 16-bit integer (uint16).
1
= ready,
0
= not ready.
Comments
•
In the keyword,
CALibration
n, the valid range of n is:
1 to 8 for
E1418A, E1418A Option 002
1 to 16 for
E1418A Option 001, E1418A Option 003.
Values of n outside these ranges will generate an error.
•
This command is used together with the
CALibration
n
:VALue:CURRent? <
value
>
query to perform current mode calibration.
•
The calibration is performed either at the CAL output terminals or at an individual
channel. The choice is made using the
DIAGnostic:CALibration:OUTPut[:MODE]
command.
–
When the
DIAG:CAL:OUTP:MODE CALBus
command (also the reset state)
is used, the output relay routes the calibration current to the CAL output
terminals (and opens all channel output terminals).
–
When the
DIAG:CAL:OUTP:MODE CHANnel
command is used, the output relay
routes the calibration current to the channel terminals (the CAL terminals are
not
connected).
The
DIAG:CAL:OUTP:MODE
command must be sent to the module before the
CALibration
n
:CONFigure:CURRent?
query.
•
This query will generate an error if the trigger system is waiting for a trigger event
when this command is received.
•
This query returns a
0
to indicate that the channel is not ready for calibration, and
a
1
to indicate the channel is ready for calibration.
•
This query is affected by the state of the program jumper for this channel on the
terminal module. To use this command:
–
the P/J jumper must be set to program selectable or, if set to jumper selectable,
–
the V/I jumper must be set to the current mode. The jumpers are described on
page 31.
•
Refer to Appendix D, “Voltage/Current Output Adjustment”, starting on page 157,
for additional information about this command.
•
Related Commands:
CALibration
n
:VALue:CURRent?
,
DIAGnostic:CALibration:OUTPut[:MODE]
CALibration
Chapter 3
HP E1418A SCPI Command Reference 70
Содержание E1418A
Страница 6: ...Notes 6 Contents HP E1418A 8 16 Channel D A Converter Module ...
Страница 10: ...Notes 10 HP E1418A User s Manual ...
Страница 12: ...12 HP E1418A User s Manual ...
Страница 105: ...TRIGger 105 HP E1418A SCPI Command Reference Chapter 3 ...
Страница 111: ...Notes HP E1418A Command Quick Reference 111 HP E1418A SCPI Command Reference Chapter 3 ...
Страница 135: ...135 HP E1418A Register Based Programming Appendix B ...
Страница 157: ...Notes 156 HP E1418A Error Messages Appendix C ...
Страница 164: ...Figure D 4 Resistance Calibration Connections CAL Appendix D Voltage Current Output Adjustment 163 ...
Страница 170: ...Notes Appendix D Voltage Current Output Adjustment 169 ...
Страница 174: ...Figure E 1 8 Channel Disassembly 172 Configuration and Disassembly Appendix E ...
Страница 175: ...Figure E 2 16 Channel Disassembly Appendix E Configuration and Disassembly 173 ...
Страница 177: ...Figure E 3 Plug on Channel Locations Figure E 4 Installing Plug on Modules Appendix E Configuration and Disassembly 175 ...
Страница 192: ...Notes 192 HP E1418A 8 16 Channel D A Converter Module Index ...