APPLy
The
APPLy
subsystem enables and outputs a current or voltage on the specified channel.
Subsystem Syntax
APPLy
n
:CURRent <
value
| MIN | MAX | DEF>
:VOLTage <
value
| MIN | MAX | DEF>
APPLy
n
:CURRent
APPLy
n
:CURRent <
value
| MIN | MAX | DEF>
outputs a current on the specified
channel, n.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
APPLy
n
keyword
1 to 8 or 1 to 16
1
<
value
>
numeric (float 32) or
discrete
-0.02 Amps to +0.02 Amps or
MIN | MAX | DEF
0
Comments
•
In the keyword,
APPLy
n, the valid range of n is:
1 to 8 for
E1418A, E1418A Option 002
1 to 16 for
E1418A Option 001, E1418A Option 003.
Values of n outside these ranges will generate an error.
•
The
<
value
>
parameter can range from -0.02 to +0.02 Amps.
MIN
sets -0.02 Amps output.
MAX
sets +0.02 Amps output.
DEF
sets 0 Amps output.
•
This command is affected by the state of the P/J jumper for this channel on the
terminal module. To use the
APPLy
n
:CURRent command:
–
the P/J jumper must be set to program selectable or, if set to jumper selectable,
–
the V/I jumper must be set to the current mode.
These jumpers are described on page 31.
•
Equivalent Commands:
SOURce
n
:FUNCtion:MODE CURR
,
SOURce
n
:CURRent[:LEVel][:IMMediate][:AMPLitude]
,
OUTPut
n
[:STATe] ON
•
*RST Condition:
–
If the P/J jumper is set to program selectable:
output mode is set to voltage, 0 volts output, output relay open.
–
If the P/J jumper is set to jumper selectable:
output mode is set by the V/I jumper, 0 volts or amps, output relay open.
APPLy
63 HP E1418A SCPI Command Reference
Chapter 3
Содержание E1418A
Страница 6: ...Notes 6 Contents HP E1418A 8 16 Channel D A Converter Module ...
Страница 10: ...Notes 10 HP E1418A User s Manual ...
Страница 12: ...12 HP E1418A User s Manual ...
Страница 105: ...TRIGger 105 HP E1418A SCPI Command Reference Chapter 3 ...
Страница 111: ...Notes HP E1418A Command Quick Reference 111 HP E1418A SCPI Command Reference Chapter 3 ...
Страница 135: ...135 HP E1418A Register Based Programming Appendix B ...
Страница 157: ...Notes 156 HP E1418A Error Messages Appendix C ...
Страница 164: ...Figure D 4 Resistance Calibration Connections CAL Appendix D Voltage Current Output Adjustment 163 ...
Страница 170: ...Notes Appendix D Voltage Current Output Adjustment 169 ...
Страница 174: ...Figure E 1 8 Channel Disassembly 172 Configuration and Disassembly Appendix E ...
Страница 175: ...Figure E 2 16 Channel Disassembly Appendix E Configuration and Disassembly 173 ...
Страница 177: ...Figure E 3 Plug on Channel Locations Figure E 4 Installing Plug on Modules Appendix E Configuration and Disassembly 175 ...
Страница 192: ...Notes 192 HP E1418A 8 16 Channel D A Converter Module Index ...