Accuracy (10 mA range) =
±
[20.0 ppm of R 5.0 ppm of Range]
Temp Coeff =
±
[2.0 ppm of R 1.0 ppm of Range]
Absolute Accuracy Factor =
±
[5.0 ppm of Reading]
For the 100 mA range:
Accuracy (100 mA range) =
±
[35.0 ppm of R 5.0 ppm Range]
Temp Coeff =
±
[2.0 ppm of R 1.0 ppm of Range]
Absolute Accuracy Factor =
±
[5.0 ppm of Reading]
Example: Calculate DCI Measurement Uncertainty
For a 4.0 mA dc reading on the 10 mA range for the conditions specified in
this manual:
Measurement Uncertainty (Amps) =
±
[10
-6
x (27.0 x R 6.0 x
Range)] =
±
[10
-6
x (27.0 x .004 + 6.0 x .010)] =
±
0.168 x 10
-6
A =
±
0.168
µ
A.
For a 16.0 mA dc reading on the 100 mA range for the conditions specified
in this manual:
Measurement Uncertainty (Amps) =
±
[10
-6
x (42.0 x R 6.0 x
Range)] =
±
[10
-6
x (42.0 x .016 + 6.0 x .100)] =
±
1.272 x 10
-6
A=
±
1.272
µ
A.
Test Accuracy
Ratio (T A R)
Calculations
For the HP E1328A D/A Converter:
Example: Calculate Voltage Output Test Accuracy Ratio
Meas Uncertainty (A) =
±
[27.0 ppm of R 6.0 pp of Range]
Reading and Range in Amps DC
Meas Uncertainty (A) =
±
[42.0 ppm of R 6.0 ppm of Range]
Reading and Range in Amps DC
Test Accuracy Ratio (TAR) =
High Limit
−
Output
Measurement Uncertainty
58 Calculating D/A Converter Accuracy
Appendix A