Table 1-2 lists the test equipment recommended for testing, adjusting, and
servicing the HP E1328A DAC. Essential requirements for each piece of
test equipment are described in the Requirements column.
Table 1-2. Recommended Test Equipment
Instrument
Requirements
Recommended
Model
Use*
Controller, HP-IB
HP-IB compatibility as defined by IEEE
Standard 488-1978 and the identical
ANSI Standard MC1.1: SH1, AH1, T2,
TE0, L2, LE0, SR0, RL0, PP0, DC0, DT0,
and C1, 2, 3, 4, 5
HP 9000 Series 300
or
IBM compatible PC with
HP BASIC
A,F,O,
P,T
Mainframe
Compatible with DAC
HP E1300A, E1301A,
E1302A, or E1401B/T,
E1421A (requires HP
E1306A)
A,F,O,
P,T
DC Standard
Voltage Range: -20.0 V to +20.0 V
Datron 4708 with
Option 10
F,O,P
Digital Multimeter
Voltage Range:
±
10 VDC
Current Range:
±
20 mA DC
Accuracy:
±
(.015% r 1 mV)
±
(.02% r 1
µ
A)
HP 3458A multimeter
or
HP E1326B multimeter
A,F,O,
P,T
*F = Functional Verification Tests, O = Operation Verification Tests, P = Performance Verification Tests,
A = Adjustments, T = Troubleshoooting
14 General Information