DAC
Description
The HP E1328A DAC is an "instrument" in the slots of a VXIbus
mainframe. As such, it is assigned an error queue, input and output buffers,
status registers, and is allocated a portion of mainframe memory for reading
storage.
NOTE
Instruments are based on the logical addresses of the plug-in modules. See
the HP 75000 Series B Installation and Getting Started Guide to set the
addresses to create an instrument.
There are two E1328A DAC functions (see Table 1-1):
•
Isolated Voltage DACs
•
Isolated Current DACs
Table 1-1. HP E1328A 4-Channel D/A Converter Functions
Function/
Feature
Chs
Description
Voltage
DAC
1,2,3,4
Output DC voltage up to
±
10.92 V independently on
each channel. Can use remote voltage sensing with
"no-fault" operation. Connect multiple channels in
series to increase the output to
±
48 Vdc for a single
module. All four channels are independently isolated
and may be floated up to 350 Vdc from ground. Each
channel can be calibrated electronically to 24-hour
specifications.
Current
DAC
1,2,3,4
Output DC current up to
±
21.8 mA independently on
each channel. Connect multiple channels in parallel
to increase the output to
±
96 mA for a single module.
All four channels are independently isolated and may
be floated up to 350 Vdc from ground. Each channel
can be calibrated electronically to 24-hour
specifications.
DAC Specifications
DAC specifications are listed in Appendix A of the HP E1328A Users
Manual. These specificaitions are the performance standards or limits
against which the instrument may be tested.
12 General Information
Chapter 1