![HP 8920B Скачать руководство пользователя страница 331](http://html.mh-extra.com/html/hp/8920b/8920b_programmers-manual_166189331.webp)
331
Chapter 5, Advanced Operations
Instrument Initialization
Device Clear (DCL) HP-IB Bus Command
The Device Clear (DCL) Reset is accomplished by sending the DCL message to
the Test Set through the HP-IB bus.
The DCL command clears the Input Buffer and Output Queue, clears any
commands in process, puts the Test Set into the Operation Complete idle state,
and prepares the Test Set to receive new commands. The DCL bus command does
not change any settings or stored data (except as noted previously), interrupt front
panel I/O, interrupt any Test Set operation in progress (except as noted
previously), or change the contents of the Status Byte Register (other than
clearing the MAV bit as a consequence of clearing the Output Queue).
The DCL bus command has no effect on the I/O CONFIGURE, CONFIGURE,
PRINT CONFIGURE, or TESTS (Main Menu, Execution Conditions, External
Devices, Printer Setup) screens.
Other operational characteristics are also affected by the DCL bus command as
follows:
•
The Power-up self-test diagnostics are not performed.
•
The HP-IB interface is not reset (any pending Service Request is not cleared)
•
Measurement triggering is not affected.
•
Calibration data is not affected.
•
The Contents of the SAVE/RECALL registers are not affected.
•
All Enable registers are unaffected: Service Request, Standard Event, Hardware #1,
Hardware #2, Operation, Calibration,Call Processing and Questionable Data/Signal.
•
All Negative Transition Filter registers are unaffected: Hardware #1, Hardware #2, Op-
eration, Calibration,Call Processing and Questionable Data/Signal.
•
All Positive Transition Filter registers are unaffected: Hardware #1, Hardware #2, Op-
eration, Calibration,Call Processing and Questionable Data/Signal.
•
The contents of the RAM memory are unaffected.
•
The contents of the Error Queue are unaffected.
•
The state of the Test Set’s display screen is unaffected.
Содержание 8920B
Страница 10: ...10 ...
Страница 16: ...Contents 16 9 Error Messages Index 673 ...
Страница 17: ...17 1 Using HP IB ...
Страница 48: ...48 Chapter 1 Using HP IB Remote Local Modes ...
Страница 49: ...49 2 Methods For Reading Measurement Results ...
Страница 61: ...61 3 HP IB Command Guidelines ...
Страница 84: ...84 Chapter 3 HP IB Command Guidelines Guidelines for Operation ...
Страница 85: ...85 4 HP IB Commands ...
Страница 90: ...90 Adjacent Channel Power ACP ...
Страница 116: ...116 Configure I O Configure ...
Страница 156: ...156 Call Processing ...
Страница 162: ...162 Display DISPlay space TDMA test TESTs TFReq THLP TIBasic TMAKe TPARm TPRint TSEQn TSPec Returns current screen TX ...
Страница 185: ...185 System System SYSTem ERRor Returns integer value quoted string ...
Страница 242: ...242 Common Command Descriptions ...
Страница 251: ...251 5 Advanced Operations ...
Страница 377: ...377 7 IBASIC Controller ...
Страница 458: ...458 Chapter 7 IBASIC Controller The TESTS Subsystem ...
Страница 640: ...640 Chapter 8 Programming The Call Processing Subsystem Example Programs 11280 SUBEND 11290 ...
Страница 643: ...643 9 Error Messages ...
Страница 672: ...672 ...
Страница 682: ...Index 682 ...