155
Chapter 6, Call Processing Subsystem
Using the CALL CONTROL Screen to test AMPS Authentication
11. Select
SSD Upd
in the
CC Order
field. The Access annunciator light will light while
the SSD Update order is in progress. If the mobile station responds properly on the
reverse control-channel, “
SSD Update Order Successful
” will be displayed
in the upper right-hand portion of the CALL CONTROL screen. The test set and the
mobile station now share common secret data and further call processing functions with
authentication can be tested. If the order failed, then “
SSD Update Order
Failed
” will appear on the upper right-hand portion of the CALL CONTROL screen.
Initializing Authentication without Registration
1. Ensure that the Test Set is in the active state.
See "Turn on the Test Set’s Control
for information on how to activate the Test Set.
2. Connect the mobile station to the Test Set as shown in
Mobile Station to the Test Set," on page 119
.
3. Turn on the mobile station.
4. If the CALL CONTROL screen is not displayed, select
CALL CNTL
from the
To
Screen
menu. The CALL CONTROL screen will be displayed.
5. Position the cursor on the
Display
field and select
Data
.
6. Select the upper
MS ID
field and then select
MIN2 MIN1
from the Choices: menu.
7. Select the lower
MS ID
field and enter the information in hexadecimal format.
8. Select
AUTHEN
from the To Screen menu. The AUTHENTICATION screen will be
displayed.
9. Select the A_KEY field and enter a valid A_KEY in decimal format.
10. Select the ESN field and enter the mobile station’s electronic serial number in a
hexadecimal format.
11. Position the cursor on the
Off/On
field and select
On
.
12. Select
CALL CNTL
from the
To Screen
menu. The CALL CONTROL screen will
be displayed.
13. Select
SSD Upd
from the
CC Order
field. The Access annunciator light will light
while the SSD Update order is in progress. If the mobile station responds properly on
the reverse control-channel, “
SSD Update Order Successful
” will be
displayed in the upper right-hand portion of the CALL CONTROL screen. The test set
and the mobile station now share common secret data and further call processing
functions with authentication can be tested. If the order failed, then “
SSD Update
Order Failed
” will appear on the upper right-hand portion of the CALL
CONTROL screen.
Содержание 8920A
Страница 14: ...Contents 14 2 Configuring Your Test Set General Operating Information 50 ...
Страница 17: ...Contents 17 5 AF Analyzer Screen Block Diagram 100 ...
Страница 19: ...Contents 19 7 Configure Screen Field Descriptions 254 ...
Страница 21: ...Contents 21 9 Duplex Test Screen Block Diagram 340 Field Descriptions 341 ...
Страница 23: ...Contents 23 11 Help Screen Field Descriptions 436 ...
Страница 24: ...Contents 24 12 I O Configure Screen Field Descriptions 438 ...
Страница 25: ...Contents 25 13 Message Screen Field Descriptions 444 ...
Страница 26: ...Contents 26 14 Oscilloscope Screen Field Descriptions 446 Using the Oscilloscope 452 ...
Страница 27: ...Contents 27 15 Print Configure Screen Field Descriptions 454 ...
Страница 29: ...Contents 29 17 RF Analyzer Screen Block Diagram 472 Field Descriptions 473 ...
Страница 30: ...Contents 30 18 RF Generator Screen Block Diagram 484 Field Descriptions 485 ...
Страница 31: ...Contents 31 19 RX Test Screen Block Diagram 494 Field Descriptions 495 ...
Страница 32: ...Contents 32 20 Service Screen Field Descriptions 502 ...
Страница 33: ...Contents 33 21 Spectrum Analyzer Screen Field Descriptions 506 Using the Spectrum Analyzer 518 ...
Страница 35: ...Contents 35 23 TX Test Screen Block Diagram 550 Field Descriptions 551 ...
Страница 38: ...Contents 38 Index 619 ...
Страница 39: ...39 1 Get Started ...
Страница 52: ...52 S HP892XX USRGUIDE manual configts fb Chapter 2 Configuring Your Test Set General Operating Information ...
Страница 98: ...98 S HP892XX USRGUIDE manual acpscrn fb Chapter 4 Adjacent Channel Power Screen Field Descriptions ...
Страница 99: ...99 5 AF Analyzer Screen ...
Страница 113: ...113 6 Call Processing Subsystem ...
Страница 129: ...129 Chapter 6 Call Processing Subsystem The CALL CONTROL Screen Figure 20 AMPS CALL CONTROL Screen with Meas Selected ...
Страница 252: ...252 S HP892XX USRGUIDE manual callscrn fb Chapter 6 Call Processing Subsystem CALL CONFIGURE Screen ...
Страница 268: ...268 S HP892XX USRGUIDE manual confscrn fb Chapter 7 Configure Screen Field Descriptions ...
Страница 269: ...269 8 Signaling Decoder Screen ...
Страница 338: ...338 S HP892XX USRGUIDE manual decoscrn fb Chapter 8 Signaling Decoder Screen Tone Sequence Decoder ...
Страница 339: ...339 9 Duplex Test Screen ...
Страница 353: ...353 10 Signaling Encoder AF Generator 2 ...
Страница 435: ...435 11 Help Screen ...
Страница 437: ...437 12 I O Configure Screen ...
Страница 443: ...443 13 Message Screen ...
Страница 445: ...445 14 Oscilloscope Screen ...
Страница 456: ...456 S HP892XX USRGUIDE manual prinscrn fb Chapter 15 Print Configure Screen Field Descriptions ...
Страница 471: ...471 17 RF Analyzer Screen ...
Страница 482: ...482 S HP892XX USRGUIDE manual rfanscrn fb Chapter 17 RF Analyzer Screen Field Descriptions ...
Страница 483: ...483 18 RF Generator Screen ...
Страница 492: ...492 S HP892XX USRGUIDE manual rfgscrn fb Chapter 18 RF Generator Screen Field Descriptions ...
Страница 493: ...493 19 RX Test Screen ...
Страница 504: ...504 S HP892XX USRGUIDE manual servscrn fb Chapter 20 Service Screen Field Descriptions ...
Страница 519: ...519 22 Tests Screen ...
Страница 548: ...548 S HP892XX USRGUIDE manual testscrn fb Chapter 22 Tests Screen ROM Programs ...
Страница 549: ...549 23 TX Test Screen ...
Страница 551: ...551 Chapter 23 TX Test Screen Field Descriptions Field Descriptions Figure 145 The TX Test Screen txscrn wmf ...
Страница 561: ...561 24 Connector Key and Knob Descriptions ...
Страница 585: ...585 25 Modifications Accessories Manuals Support ...
Страница 605: ...605 Error Messages ...
Страница 618: ...618 S HP892XX USRGUIDE manual errormsg fb Error Messages ...