1-8
Quick Start
Measuring Fiber Devices
Measuring Fiber Devices
In this section, you’ll learn how to optimize measurements on fiber devices.
You can, of course, skip this section and immediately begin making measure-
ments on your devices. However, be sure to read this material later to increase
your understanding of chromatic dispersion measurements.
By adjusting the RF modulation frequency, you can optimize measurement
accuracy. (Refer to the sidebar,
“Assumptions required for this optimization
.) For information on changing the modulation fre-
quency,
refer to “CD/Gain Meas command” on page 2-15
.
Because the system’s accuracy varies as a function of RF modulation fre-
quency, for optimum accuracy keep the modulation frequency above 1 GHz
and preferably at 2 GHz. This puts some restrictions on the wavelength step
size because too large of a step size results in a phase "wrap-around" where
the phase jump exceeds 360
°
as the wavelength is stepped. Use the following
formula to determine the maximum step size:
where:
(
∆λ
∆λ
∆λ
∆λ
)
MAX
is the maximum allowed step size (in nm) to keep the modulation
frequency at the desired value fm.
f
m
is the desired modulation frequency (in GHz).
S
0
is the dispersion slope in ps/(nm2×km).
λ
SPAN
is the spectral span (stop
λ
– start
λ
) of the measurement in nm.
L is the length of the system/fiber in km.
Suppose, for example, that you wanted to measure chromatic dispersion on a
70 km spool of dispersion shifted fiber between 1515 nm and 1575 nm. Select
the optimum modulation frequency of 2 GHz, and use the above equation to
determine the maximum allowed wavelength step size. Assuming the fiber’s
dispersion slope is 0.07 ps/(nm2 km), then:
5 0 0
S
o
ƒ
m
λ
S P A N
L
(∆λ)
M A X
=
Содержание 86037B
Страница 1: ...User s Guide HP 86037B Chromatic Dispersion Test System ...
Страница 31: ...2 13 Operation The Main Window Figure 2 4 Repeat Measurement Selector ...
Страница 72: ...3 16 Normalization Setting Normalization Preferences ...
Страница 94: ...4 22 Remote Control TES Parameters Main chromatic dispersion screen ...
Страница 100: ...4 28 Remote Control TES Parameters CD and Gain Measurement Settings Dialog Box ...
Страница 102: ...4 30 Remote Control TES Parameters Length Measurement Settings Dialog Box ...
Страница 104: ...4 32 Remote Control TES Parameters Normalization Settings Dialog Box ...
Страница 106: ...4 34 Remote Control TES Parameters New Normalization Button Dialog Box ...
Страница 120: ...4 48 Remote Control Output Parameters ...
Страница 158: ...5 38 Maintenance Hewlett Packard Sales and Service Offices ...
Страница 165: ...6 7 Installation Installation 2 3 4 5 1 8 7 Fan Pack 6 6 Power Distribution Assembly 1 7 Power Distribution Assembly 2 ...
Страница 170: ...6 12 Installation Installation ...
Страница 174: ...7 4 Reference Theory of Operation Figure 7 3 HP 86037B Block Diagram Options 121 and 122 ...
Страница 175: ...7 5 Reference Theory of Operation Figure 7 4 The HP 86037B Option 114 block diagram ...
Страница 190: ...7 20 Reference System Options ...
Страница 191: ...8 Specifications 8 3 Regulatory Information 8 7 Specifications and Regulatory Information ...
Страница 198: ...8 8 Specifications and Regulatory Information Regulatory Information Declaration of Conformity ...
Страница 204: ...Index 6 Index ...