Operation
Measurement System Calibration
Measurement System Calibration
A perfect vector network analyzer exhibits flat frequency response, no
impedance mismatches, and infinite isolation between channels. In an actual
system all of these characteristics are imperfect, but are generally repeatable
and predictable. To correct for these systematic errors, measurement
calibration is performed using known impedance standards connected at the
measurement ports. All of the required standards are supplied with the
HP 11644A series calibration kits. To be sure the system is meeting
measurement specifications, refer to Chapter 3, “Performance Verification”.
The HP 85106D with an HP 11644A series calibration kit can use any of the
accuracy enhancement models available with the HP 8510C. For
measurement of 2-port devices use, Full 2-Port, TRL 2-Port, or a
combination of Transmission Response and Reflection 2-Port calibrations.
Refer to the standard HP 8510C Operating and Programming manual and to
the calibration kit manual for descriptions of the calibration types.
Procedures for the TRL 2-Port, Response, Response & Isolation, and 1-Port
calibration types are described in this section.
Warm-up Time
Allow at least one hour for the system to reach a stable temperature after
power is turned on before trying to get highly repeatable measurements. In
addition, a stable ambient temperature helps to minimize time-base
variations, as well as physical dimensional changes that appear as short or
long term drift in magnitude and phase measurements.
Before Starting a
Measurement
Calibration
Check to be certain the appropriate CAL KIT definition is installed in the
HP 8510C. This definition is part of the Machine Dump file previously
loaded, and is contained on the disk in each HP 11644A series calibration
kit. Press [
CAL
] and read the {
CAL 1
}or {
CAL 2
} softkey label to confirm the
calibration kit label.
Setting the Stimulus
Values
Set all of the desired Stimulus characteristics for device measurement. These
include the frequency range or frequency list, the number of points, Source
#1 power level, and so forth. If the DUT must be measured using different
stimulus settings, then make separate measurement calibrations for each
frequency range (except where the HP 8510C Frequency Subset feature can
be used).
Содержание 85106D
Страница 11: ...x HP 85106D mm Wave System ...
Страница 95: ...4 2 HP 85106D mm Wave System Service and Troubleshooting Figure 4 1 HP 85106D System Block Diagram ...
Страница 96: ...HP 85106D mm Wave System 4 3 Service and Troubleshooting Figure 4 2 HP 85106D Option 001 System Block Diagram ...
Страница 117: ...4 24 HP 85106D mm Wave System Service and Troubleshooting Check Unratioed Power Levels ...
Страница 139: ...5 22 HP 85106D mm Wave System HP 85104A Test Set Module Troubleshooting Procedures ...
Страница 140: ......
Страница 141: ......
Страница 142: ......
Страница 143: ......
Страница 149: ...5 36 HP 85106D mm Wave System HP 85104A Test Set Module Assembly Procedure Figure 5 12 HP 85104A Assembly Procedure ...
Страница 153: ...5 40 HP 85106D mm Wave System HP 85104A Test Set Module Replaceable Parts Figure 5 13 HP Q85104A Replaceable Parts ...
Страница 156: ...HP 85106D mm Wave System 5 43 HP 85104A Test Set Module Replaceable Parts Figure 5 14 HP U85104A Replaceable Parts ...
Страница 159: ...5 46 HP 85106D mm Wave System HP 85104A Test Set Module Replaceable Parts Figure 5 15 HP V85104A Replaceable Parts ...
Страница 162: ...HP 85106D mm Wave System 5 49 HP 85104A Test Set Module Replaceable Parts Figure 5 16 HP W85104A Replaceable Parts ...
Страница 208: ......
Страница 212: ...6 48 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 28 HP 85105A Major Assemblies ...
Страница 214: ...6 50 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 29 HP 85105A Semi rigid RF Cables ...
Страница 218: ...6 54 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 31 Wire Harnesses ...
Страница 220: ...6 56 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 32 HP 85105A Front Panel ...
Страница 222: ...6 58 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 33 Rear Panel Standard Option 050 ...
Страница 228: ...6 64 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 37 HP 85105A Accessories Standard ...
Страница 238: ...B 2 HP 85106D mm Wave System Upgrade Paths ...
Страница 243: ...HP 85106D mm Wave System v work surface installation 1 12 ...
Страница 244: ... vi HP 85106D mm Wave System ...