HP 85106D mm-Wave System
5-1
5
HP 85104A Test Set Module
Introduction
This section documents the operation, troubleshooting techniques, and
replaceable parts of the HP 85104A test set module.
Theory Of Operation
A pair of HP 85104A test set modules are used to make either reflection,
transmission, or S-parameter measurements at mm-wave frequencies, in
conjunction with an HP 85105A mm-wave controller. The mm-wave
controller routes the LO and RF signals from RF and LO sources to the test
set modules. Switching from port 1 to port 2 is also performed by the
mm-wave controller. The test set modules separate the incident from the
reflected RF signal and then down convert those signals to a 20 MHz IF
frequency.
Refer to the HP 85104A block diagrams (Figure 5-7 on page 5-23 through
Figure 5-10 on page 5-29) while reading the following description. The RF
is received from the mm-wave controller and input to the source module
block of the test set module. The RF is then multiplied six times in the case
of W-band (2 times for R-band, 8 times for Q-band, 8 times for U-band and
4 times for V-band). This multiplied RF signal then passes through a
coupler/detector (only in V and W bands) where the signal is sampled for
use in an ALC loop. The RF continues through an isolator and a dual
directional coupler. A portion of the incident signal is coupled off, passed
through an isolator, and into a harmonic mixer, A11. The signal received at
the test port (either transmitted from another test set module or reflected
from a device under test) is coupled off, passed through an isolator, and into
a harmonic mixer, A12. The 2-8 GHz LO signal is input from the source
through the mm-wave controller. The signal is then divided and input to the
harmonic mixer. In the case of W-band, the sixteenth harmonic of the LO
mixes with the RF so that a 20 MHz IF frequency is output. The LO
harmonic product must be 20 MHz higher in frequency than the RF signal.
The IF frequencies (incident and reflected) are then output to the mm-wave
controller for further processing. The mixing harmonics for other bands are;
R-band: 8, Q-band: 10, U-band: 10, and V-band: 14. The isolators in the test
set modules are used to keep reflections from the dual directional coupler,
A8, from interfering with the source, and to keep reflections from the
mixers, All and A12, from interfering with the incident and reflected RF
signals.
Warranty
Refer to Chapter 1, “Getting Started” in this manual for warranty
information about the HP 85104A.
Содержание 85106D
Страница 11: ...x HP 85106D mm Wave System ...
Страница 95: ...4 2 HP 85106D mm Wave System Service and Troubleshooting Figure 4 1 HP 85106D System Block Diagram ...
Страница 96: ...HP 85106D mm Wave System 4 3 Service and Troubleshooting Figure 4 2 HP 85106D Option 001 System Block Diagram ...
Страница 117: ...4 24 HP 85106D mm Wave System Service and Troubleshooting Check Unratioed Power Levels ...
Страница 139: ...5 22 HP 85106D mm Wave System HP 85104A Test Set Module Troubleshooting Procedures ...
Страница 140: ......
Страница 141: ......
Страница 142: ......
Страница 143: ......
Страница 149: ...5 36 HP 85106D mm Wave System HP 85104A Test Set Module Assembly Procedure Figure 5 12 HP 85104A Assembly Procedure ...
Страница 153: ...5 40 HP 85106D mm Wave System HP 85104A Test Set Module Replaceable Parts Figure 5 13 HP Q85104A Replaceable Parts ...
Страница 156: ...HP 85106D mm Wave System 5 43 HP 85104A Test Set Module Replaceable Parts Figure 5 14 HP U85104A Replaceable Parts ...
Страница 159: ...5 46 HP 85106D mm Wave System HP 85104A Test Set Module Replaceable Parts Figure 5 15 HP V85104A Replaceable Parts ...
Страница 162: ...HP 85106D mm Wave System 5 49 HP 85104A Test Set Module Replaceable Parts Figure 5 16 HP W85104A Replaceable Parts ...
Страница 208: ......
Страница 212: ...6 48 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 28 HP 85105A Major Assemblies ...
Страница 214: ...6 50 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 29 HP 85105A Semi rigid RF Cables ...
Страница 218: ...6 54 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 31 Wire Harnesses ...
Страница 220: ...6 56 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 32 HP 85105A Front Panel ...
Страница 222: ...6 58 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 33 Rear Panel Standard Option 050 ...
Страница 228: ...6 64 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 37 HP 85105A Accessories Standard ...
Страница 238: ...B 2 HP 85106D mm Wave System Upgrade Paths ...
Страница 243: ...HP 85106D mm Wave System v work surface installation 1 12 ...
Страница 244: ... vi HP 85106D mm Wave System ...