75
Programming Reference
SCPI Instrument Command List
Command
:VOLT[1]:LIM:STAT
Long
[:SOURce]:VOLTage[1]:LIMit:STATe
Form
Set & Query
Parameter
ON | OFF | 1 | 0
*RST value
OFF
Description
Use this command to switch the output limits on or off. When you switch
on the output limits, you cannot program the output levels beyond the
programmed limits, until you switch off the voltage limits. The limits
apply whether you program high/low levels or amplitude/offset levels.
N O T E
You can switch the limits on and off in both the
[:SOURce]:CURRent
and the
[:SOURce]:VOLTage
subsystems
but the current and voltage
limits are not enabled/disabled independently
. The voltage and current
limits are always enabled/disabled together.
Example
To set and activate the high and low voltage limits:
Command
:STATus:OPERation
This command tree accesses the
OPERation
status group.
The
OPERation
status group is not used by the instrument. Therefore, this
command tree is redundant
.
•
:STATus:OPERation[:EVENt]?
•
:STATus:OPERation:CONDition?
•
:STATus:OPERation:ENABle
•
:STATus:OPERation:NTRansition
•
:STATus:OPERation:PTRansition
:HOLD VOLT
:VOLT:LIM 3V
:VOLT:LIM:LOW 0V
:VOLT:LIM:STAT ON
Enable VOLTAGE subsystem
Set OUTPUT high level voltage limit to 3 V
Set OUTPUT low level voltage limit to 0 V
Switch on OUTPUT limits
Содержание 81101A
Страница 1: ...H HP 81101A 50 MHz Pulse Generator Reference Guide ...
Страница 10: ...10 Safety Information ...
Страница 88: ...88 Programming Reference SCPI Instrument Command List ...
Страница 112: ...112 Specifications Pulse Parameter Definitions ...
Страница 116: ...116 Index ...
Страница 118: ...Hewlett Packard GmbH Copyright 1998 Edition E0998 Printed in Germany September 1998 BAR CODE to go here 81101 91011 ...