Troubleshooting the MSD
Mass spectral symptoms
77
Relative abundance of
m/z
502 is less than 3%
Autotune should give an
m/z
502 relative abundance greater than 3%. The rela-
tive abundance of
m/z
502 can, however, vary a great deal depending on column
flow, ion source temperature, and other variables. As long as relative abundance
is above 3%, the stability of the relative abundance is more important than the
absolute value. If you observe significant changes in the relative abundance of
m/z
502 for a fixed set of operating parameters, there may be a problem. The
charts in the MSD ChemStation software are useful for identifying changes. Select
View Tunes
from the Qualify menu in the Instrument Control view.
Low
relative
abundance of
m/z
502 should not be confused with low
absolute
abundances at high masses. Sensitivity at high masses can be excellent even if the
relative abundance of
m/z
502 is near 3%. If your MSD produces low absolute
abundances at high masses, refer to the symptom
High mass sensitivity is poor
.
Tune programs other than autotune have different relative abundance targets. The
DFTPP and BFB target tune programs tune the HP 5973 MSD to achieve about a
0.8% ratio of
m/z
502/69.
•
Tune program/tune file has a different relative abundance target
(3% only applies to Autotune)
•
Not enough time for the MSD to warm up and pump down
•
Vacuum manifold pressure is too high
•
Ion source temperature is too high
•
Column (carrier gas) flow is too high
•
Poor filament operation
•
Dirty ion source
•
Air leak
•
Incorrect dc polarity on the quadrupole mass filter
Spectra look different from those acquired with other HP MSDs
Ion ratios are different from those in older HP MSDs. This is due to the HED detec-
tor, and is normal. To get spectra similar to older MSDs, use
Standard Spectra Tune
,
available in the Manual Tune view.
Содержание 5973
Страница 1: ...H Hardware Manual HP 5973 Mass Selective Detector ...
Страница 3: ...HP 5973 Mass Selective Detector Hardware Manual ...
Страница 25: ...1 Installing GC Columns To condition a capillary column 25 ...
Страница 30: ...30 ...
Страница 37: ...2 Operating the MSD To view MSD analyzer temperature and vacuum status 37 ...
Страница 39: ...2 Operating the MSD To set monitors for MSD temperature and vacuum status 39 ...
Страница 41: ...2 Operating the MSD To set the MSD analyzer temperatures 41 ...
Страница 43: ...2 Operating the MSD To set the interface temperature from the ChemStation 43 ...
Страница 51: ...2 Operating the MSD To remove the MSD covers 51 Upper cover Lower cover ...
Страница 92: ...92 ...
Страница 103: ...4 Maintaining the MSD To check and add foreline pump oil 103 Lower line Oil level window Fill cap Upper line ...
Страница 105: ...4 Maintaining the MSD To drain the foreline pump 105 Fill cap Drain plug Pump motor ...
Страница 107: ...4 Maintaining the MSD To refill the foreline pump 107 Fill cap Upper line Oil level window Drain plug ...
Страница 109: ...4 Maintaining the MSD To replace the oil trap 109 Oil trap Clamp O ring assembly Oil trap adapter Foreline pump ...
Страница 111: ...4 Maintaining the MSD To check the diffusion pump fluid 111 Fan Fluid level window sight glass ...
Страница 113: ...4 Maintaining the MSD To separate the MSD from the GC 113 ...
Страница 121: ...4 Maintaining the MSD To reconnect the MSD to the GC 121 ...
Страница 123: ...4 Maintaining the MSD To remove the calibration vial 123 Calibration valve Collar Calibration vial ...
Страница 125: ...4 Maintaining the MSD To refill and reinstall the calibration vial 125 Calibration valve Collar Calibration vial 6 mm ...
Страница 141: ...4 Maintaining the MSD To lubricate the side plate O ring 141 Side plate O ring ...
Страница 159: ...4 Maintaining the MSD To remove a filament 159 Filament 2 Filament 1 ...
Страница 161: ...4 Maintaining the MSD To reinstall a filament 161 Filament 2 Filament 1 ...
Страница 210: ...210 ...
Страница 211: ...6 GC MSD Interface This chapter describes the function of the GC MSD interface GC MSD interface 212 ...
Страница 214: ...214 ...
Страница 232: ...232 7 Analyzer Analyzer heaters and radiators ...
Страница 280: ...280 ...
Страница 294: ...294 Index ...
Страница 296: ...294 Chapter ...
Страница 297: ...H Manual Part Number G1099 90027 Copyright 1998 Hewlett Packard Printed in USA 11 98 ...