Maintenance
7RUHILOOWKH&,FDOLEUDQWYLDO
To refill the CI calibrant vial
Materials needed:
PFDTD calibrant (8500-8130)
1
Set the reagent gas flow to
Gas Off
2
Vent the MSD
3
Remove the capillary column from the GC/MSD interface
4
Pull the MSD away from the GC
5
Loosen the nut holding the vial in place
6
Remove the calibrant vial
7
Fill the vial to within 0.5 cm of the top with fresh PFDTD calibrant
(8500-8130).
8
Replace the vial and tighten the nut.
9
Reposition the MSD next to the GC.
10
Reinstall the capillary column.
11
Pump down the MSD. See page 40.
12
Purge the calibration valve. See page 99.
C A U T I O N
Do
not
rinse the vial with any solvents.
Never
expose the inside of the vial to chlorinated
solvents or isopropyl alcohol or water — this will result in severe loss of CI sensitivity.
C A U T I O N
After removing the calibrant vial, you
must
purge the calibration valve. Failure to do so
will result in severe contamination of the ion source and damage to the electron multiplier
Содержание 5973 MSD
Страница 1: ...H Hardware Manual HP 5973 MSD PCI NCI ...
Страница 6: ...6 ...
Страница 7: ...1 General Information Specifications 9 Reagent gases and regulators 10 ...
Страница 13: ... Hardware Reference 13 ...
Страница 25: ... Chemical Ionization Theory 3RVLWLYH WKHRU 25 ...
Страница 31: ... Chemical Ionization Theory 1HJDWLYH WKHRU 31 ...
Страница 34: ...34 ...
Страница 92: ...92 ...
Страница 103: ...8 Parts GC MSD interface parts 106 CI ion source parts 108 Flow control module parts 110 Miscellaneous parts 113 ...
Страница 114: ...114 ...
Страница 120: ...120 ...
Страница 122: ...H Manual Part Number G1999 90001 Copyright 1998 Hewlett Packard Printed in USA 2 98 ...