Troubleshooting
([FHVVLYHQRLVHRUORZVLJQDOWRQRLVHUDWLR
Excessive noise or low signal-to-noise ratio
Possible Cause
The GC injection port needs maintenance.
Action
Refer to the HP 6890 GC manual.
Possible Cause
The CI ion source is dirty.
Action
Clean the ion source. See the
Maintenance
chapter in this manual for more
information.
Possible Cause
Poor vacuum
Action
Check the pressure on the high vacuum gauge controller.
Possible Cause
Air leak.
Action
Run Methane Pre-tune (in PCI). Large peak at
m/z
32 indicates air in the system.
Check for and correct any leaks. See the
Leaks
section at the beginning of this
chapter.
Possible Cause
Saturated methane / isobutane gas purifier.
Action
Replace the gas purifier.