CI setup
&,VHWXS
37
CI setup
In setting up your CI MD for operation in CI mode requires special care to avoid
contamination and air leaks.
General guidelines
•
Before venting in EI mode, verify that the GC/MSD system is performing
correctly.
•
Make sure the carrier gas plumbing and GC have no air leaks. (EI mode.)
•
Make sure the MSD vacuum system has no air leaks.
•
Make sure the reagent gas inlet line(s) are equipped with gas purifiers (not
applicable for ammonia.)
•
Use extra-high purity reagent gases; 99.97% or better for methane and as pure
as is available for other reagent gases.